Background
Dynamic hip screws (DHS) and proximal femoral nail anti-rotation (PFNA) were recommended for basicervical femoral neck fracture (BFNF), however, with high rate of postoperative femoral neck shortening. The proximal femoral bionic nail (PFBN) was designed to decrease the postoperative complications associated with DHS and PFNA. The aim of this study is to compare the biomechanical characters of DHS, PFNA, and PFBN for fixation of BFNF.
Methods
Using finite element analysis, we created a three-dimensional model of the BFNF for this investigation. The PFBN group, the PFNA group and the DHS + DS group were our three test groups. For each fracture group, the von Mises stress and displacements of the femur and internal fixation components were measured under 2100 N axial loads.
Results
The PFBN group demonstrated the lowest stress on the implants, significantly lower than the PFNA and DHS + DS groups. In terms of stress on the implants, the PFBN group exhibited the best performance, with the lowest stress concentration at 112.0 MPa, followed by the PFNA group at 124.8 MPa and the DHS + DS group at 149.8 MPa. The PFBA group demonstrated the smallest displacement at the fracture interface, measuring 0.21 mm, coupled with a fracture interface pressure of 17.41 MPa, signifying excellent stability.
Conclusions
Compared with DHS and PFNA, PFBN has advantages in stress distribution and biological stability. We believe the concept of triangle fixation will be helpful to reduce femoral neck shortening associated with DHS and PFNA and thus improve the prognosis of BFNF.
Funder
Health Commission of Hebei Province
National Natural Science Foundation of China
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献