Author:
Zhou Quan-kun,Zeng Fan-hui,Tu Jian-long,Dong Zhang-qing,Ding Zhi-Hui
Abstract
Abstract
Purpose
To compare the effect of conventional pedicle screw (CPS) and cement-augmented pedicle screw instrumentation (CAPSI) on adjacent segment degeneration (ASD).
Methods
A normal male volunteer without a history of spinal disease was selected, lumbar CT data was collected, an intact L3-S1 three-dimensional finite element model was created by software including Mimics, Geomagic, and SolidWorks, and the fixation methods were performed accordingly. A common pedicle screw model and a cement-augmented pedicle screw model of L4–L5 with fusion and internal fixation were constructed. With ANSYS Workbench 17.0, a 500 N load was applied to the upper surface of L3 to simulate the weight of a human body, and a 7.5 N m moment was applied at the neutral point to simulate flexion, extension, left/right bending, left/right rotation of the spine. The peak von Mises stress of intervertebral disc and the range of motion (ROM) on the adjacent segments (L3–4 and L5–S1) were compared.
Results
The validity of the intact model shows that the ROM of the model is similar to that of a cadaveric study. Compared with the intact model, CPS model and CAPSI model in all motion patterns increased the ROM of adjacent segments. The intervertebral disc stress and the ROM of adjacent segments were found to be higher in the CAPSI model than in the CPS model, especially in L3–4.
Conclusion
In general, the biomechanical analysis of an osteoporotic lumbar spine showed that both CPS and CAPSI can increase the ROM and disc stresses of osteoporotic lumbar models, and compared with CPS, CAPSI is more likely to increase the potential risk of adjacent segment degeneration.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Reference34 articles.
1. Hu L, Wu T, Liu H, et al. Influence of fusion on the behavior of adjacent disc arthroplasty in contiguous 2-level hybrid surgery in vivo. World Neurosurg. 2019.
2. Korovessis P, Syrimpeis V, Tsekouras V, Vardakastanis K, Fennema P. PEEK versus Silicon Interspinous Spacer for Reduction of Supradjacent Segment Degeneration following Decompression and Short-Segment Instrumentation for Degenerative Lumbar Spinal Stenosis. Adv Orthop. 2018;2018:1623647.
3. Wang W, Liu C, Li J, et al. Comparison of the fenestrated pedicle screw and conventional pedicle screw in minimally percutaneous fixation for the treatment of spondylolisthesis with osteoporotic spine. Clin Neurol Neurosurg. 2019;183:105377.
4. Mo GY, Guo HZ, Guo DQ, et al. Augmented pedicle trajectory applied on the osteoporotic spine with lumbar degenerative disease: mid-term outcome. J Orthop Surg Res. 2019;14(1):170.
5. Guo HZ, Tang YC, Guo DQ, et al. The cement leakage in cement-augmented pedicle screw instrumentation in degenerative lumbosacral diseases: a retrospective analysis of 202 cases and 950 augmented pedicle screws. Eur Spine J. 2019;28(7):1661–9.