Abstract
Abstract
Background
This study aimed to evaluate the efficacy and safety of Scepter dual-lumen balloon catheter for transarterial Onyx embolization of dural arteriovenous fistula (DAVF).
Methods
Transarterial Onyx embolization using a Scepter dual-lumen balloon catheter (Scepter-assisted Onyx embolization) for DAVF was attempted in a total of 35 patients (mean age, 52.5 years; M:F = 24:11) between October 2012 and December 2018. The results of Scepter-assisted Onyx embolization were evaluated with respect to total procedural and Onyx injection times, the types and number of feeders requiring embolization, angiographic and clinical outcomes, and treatment-related complications.
Results
Initial presentations were non-hemorrhagic neurological deficits in 10, intracranial hemorrhage in 8, seizure in 7, headache in 7, and intractable tinnitus in 3. All DAVF were aggressive type (Borden type 2, 14.3 %; type 3, 85.7 %). Scepter-assisted Onyx embolization resulted in immediately complete occlusion in 33 patients (94.3 %) and near complete occlusion in 2 patients. Middle meningeal artery (51.4 %) was the most commonly used for Scepter-assisted technique, followed by occipital artery (42.9 %), ascending pharyngeal artery (2.9 %) and superficial temporal artery (2.9 %). There was no difference in complete occlusion rate between middle meningeal artery and the other arteries (94.4 % versus 94.1 %). The median number of total feeders embolized was 1 (range, 1–3). The median total procedural time was 45 minutes (range, 21 minutes – 127 minutes) and the median Onyx injection time was 11 minutes (range, 3 minutes – 25 minutes). All patients recovered completely (n = 31) or partially (n = 4) from presenting symptoms. Treatment-related complications occurred in 2 patients, of whom one had a permanent morbidity (2.8 %, ipsilateral facial nerve palsy). No patient showed a recurrence on follow-up imaging (median, 15 months; range, 3–56 months).
Conclusions
Scepter-assisted transarterial Onyx embolization showed a very high complete occlusion rate with a low morbidity and no recurrence in aggressive type DAVF. Scepter dual-lumen balloon catheter seems to be a useful tool for transarterial Onyx embolization of DAVF.
Publisher
Springer Science and Business Media LLC
Subject
Neurology (clinical),General Medicine
Reference15 articles.
1. Kiyosue H, Hori Y, Okahara M, Tanoue S, Sagara Y, Matsumoto S, et al. Treatment of intracranial dural arteriovenous fistulas: current strategies based on location and hemodynamics, and alternative techniques of transcatheter embolization. Radiographics. 2004;24:1637–53.
2. Starke RM, McCarthy DJ, Chen CJ, Kano H, McShane B, Lee J, et al. Evaluation of stereotactic radiosurgery for cerebral dural arteriovenous fistulas in a multicenter international consortium. J Neurosurg. 2019;1:1–8.
3. Jagadeesan BD, Grigoryan M, Hassan AE, Grande AW, Tummala RP. Endovascular balloon-assisted embolization of intracranial and cervical arteriovenous malformations using dual-lumen coaxial balloon microcatheters and Onyx: initial experience. Neurosurgery. 2013;73:238–43.
4. Kim JW, Kim BM, Park KY, Kim DJ, Kim DI. Onyx Embolization for Isolated Type Dural Arteriovenous Fistula Using a Dual-Lumen Balloon Catheter. Neurosurgery. 2016;78:627–36.