Author:
King Shane T.,Eveld Maura E.,Zelik Karl E.,Goldfarb Michael
Abstract
Abstract
Background
Falls due to stumbling are prevalent for transfemoral prosthesis users and may lead to increased injury risk. This preliminary case series analyzes the transfemoral prosthesis user stumble recovery response to highlight key deficits in current commercially-available prostheses and proposes potential interventions to improve recovery outcomes.
Methods
Six transfemoral prosthesis users were perturbed on their prosthetic limb at least three times while walking on a treadmill using obstacle perturbations in early, mid and late swing. Kinematic data were collected to characterize the response, while fall rate and key kinematic recovery metrics were used to assess the quality of recovery and highlight functional deficits in current commercially-available prostheses.
Results
Across all participants, 13 (54%) of the 24 trials resulted in a fall (defined as > 50% body-weight support) with all but one participant (83%) falling at least once and two participants (33%) falling every time. In contrast, in a previous study of seven young, unimpaired, non-prosthesis users using the same experimental apparatus, no falls occurred across 190 trials. For the transfemoral prosthesis users, early swing had the highest rate of falling at 64%, followed by mid-swing at 57%, and then late swing at 33%. The trend in falls was mirrored by the kinematic recovery metrics (peak trunk angle, peak trunk angular velocity, forward reach of the perturbed limb, and knee angle at ground contact). In early swing all four metrics were deficient compared to non-prosthesis user controls. In mid swing, all but trunk angular velocity were deficient. In late swing only forward reach was deficient.
Conclusion
Based on the stumble recovery responses, four potential deficiencies were identified in the response of the knee prostheses: (1) insufficient resistance to stance knee flexion upon ground contact; (2) insufficient swing extension after a perturbation; (3) difficulty initiating swing flexion following a perturbation; and (4) excessive impedance against swing flexion in early swing preventing the potential utilization of the elevating strategy. Each of these issues can potentially be addressed by mechanical or mechatronic changes to prosthetic design to improve quality of recovery and reduce the likelihood a fall.
Funder
National Institutes of Health
Publisher
Springer Science and Business Media LLC
Reference36 articles.
1. Verma SK, Willetts JL, Corns HL, Marucci-Wellman HR, Lombardi DA, Courtney TK. Falls and fall-related injuries among community-dwelling adults in the United States. PLoS ONE. 2016;11(3):0150939.
2. Talbot LA, Musiol RJ, Witham EK, Metter EJ. Falls in young, middle-aged and older community dwelling adults: perceived cause, environmental factors and injury. BMC Public Health. 2005;5:1–9.
3. Tinetti ME, Baker DI, McAvay G, Claus EB, Garrett P, Gottschalk M, Koch ML, Trainor K, Horwitz RI. A multifactorial intervention to reduce the risk of falling among elderly people living in the community. N Engl J Med. 1994;331(13):821–7.
4. Leamon TB, Murphy PL. Occupational slips and falls: more than a trivial problem. Ergonomics. 1995;38(3):487–98.
5. Hunter SW, Batchelor F, Hill KD, Hill A-M, Mackintosh S, Payne M. Risk factors for falls in people with a lower limb amputation: a systematic review. Pm &r. 2017;9(2):170–80.