Author:
Khantan Mehdi,Avery Mikael,Aung Phyo Thuta,Zarin Rachel M.,Hammelef Emma,Shawki Nabila,Serruya Mijail Demian,Napoli Alessandro
Abstract
Abstract
Background
Active upper extremity (UE) assistive devices have the potential to restore independent functional movement in individuals with UE impairment due to neuromuscular diseases or injury-induced chronic weakness. Academically fabricated UE assistive devices are not usually optimized for activities of daily living (ADLs), whereas commercially available alternatives tend to lack flexibility in control and activation methods. Both options are typically difficult to don and doff and may be uncomfortable for extensive daily use due to their lack of personalization. To overcome these limitations, we have designed, developed, and clinically evaluated the NuroSleeve, an innovative user-centered UE hybrid orthosis.
Methods
This study introduces the design, implementation, and clinical evaluation of the NuroSleeve, a user-centered hybrid device that incorporates a lightweight, easy to don and doff 3D-printed motorized UE orthosis and a functional electrical stimulation (FES) component. Our primary goals are to develop a customized hybrid device that individuals with UE neuromuscular impairment can use to perform ADLs and to evaluate the benefits of incorporating the device into occupational therapy sessions. The trial is designed as a prospective, open-label, single-cohort feasibility study of eight-week sessions combined with at-home use of the device and implements an iterative device design process where feedback from participants and therapists informs design improvement cycles.
Results
All participants learned how to independently don, doff, and use the NuroSleeve in ADLs, both in clinical therapy and in their home environments. All participants showed improvements in their Canadian Occupational Performance Measure (COPM), which was the primary clinical trial outcome measure. Furthermore, participants and therapists provided valuable feedback to guide further development.
Conclusions
Our results from non-clinical testing and clinical evaluation demonstrate that the NuroSleeve has met feasibility and safety goals and effectively improved independent voluntary function during ADLs. The study’s encouraging preliminary findings indicate that the NuroSleeve has met its technical and clinical objectives while improving upon the limitations of the existing UE orthoses owing to its personalized and flexible approach to hardware and firmware design.
Trial Registration: ClinicalTrials.gov identifier: NCT04798378, https://clinicaltrials.gov/ct2/show/NCT04798378, date of registration: March 15, 2021.
Publisher
Springer Science and Business Media LLC
Subject
Health Informatics,Rehabilitation
Reference87 articles.
1. O’Neill C, Proietti T, Nuckols K, Clarke ME, Hohimer CJ, Cloutier A, et al. Inflatable soft wearable robot for reducing therapist fatigue during upper extremity rehabilitation in severe stroke. IEEE Robot Autom Lett. 2020;5:3899–906.
2. Stroke Facts & Statistics. Stroke Awareness Foundation. 2021. https://www.strokeinfo.org/stroke-facts-statistics/. Accessed 17 Dec 2022.
3. Allison R, Shenton L, Bamforth K, Kilbride C, Richards D. Incidence, time course and predictors of impairments relating to caring for the profoundly affected arm after stroke: a systematic review. Physiother Res Int. 2016;21:210–27.
4. Dunkelberger N, Schearer EM, O’Malley MK. A review of methods for achieving upper limb movement following spinal cord injury through hybrid muscle stimulation and robotic assistance. Exp Neurol. 2020;328: 113274.
5. What is Muscular Dystrophy? Centers for Disease Control and Prevention. 2020. https://www.cdc.gov/ncbddd/musculardystrophy/facts.html. Accessed 4 Nov 2022.
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献