Author:
Ecker Hannes,Kolvenbach Simone,Stranz Sebastian,Herff Holger,Wetsch Wolfgang A.
Abstract
Abstract
Background
Endotracheal intubation continues to be the gold standard for securing the airway in emergency situations. Difficult intubation is still a dreadful situation when securing the airway.
Objective
To compare VieScope with Glidescope and conventional Macintosh laryngoscopy (MAC) in a simulated difficult airway situation.
Methods
In this randomized controlled simulation trial, 35 anesthesiologists performed endotracheal intubation using VieScope, GlideScope and MAC in a randomized order on a certified airway manikin with difficult airway.
Results
For the primary endpoint of correct tube position, no statistical difference was found (p = 0.137). Time until intubation for GlideScope (27.5 ± 20.3 s) and MAC (20.8 ± 8.1 s) were shorter compared to the VieScope (36.3 ± 10.1 s). Time to first ventilation, GlideScope (39.3 ± 21.6 s) and MAC (31.9 ± 9.5 s) were also shorter compared to the VieScope (46.5 ± 12.4 s). There was no difference shown between handling time for VieScope (20.7 ± 7.0 s) and time until intubation with GlideScope or MAC. Participants stated a better Cormack & Lehane Score with VieScope, compared to direct laryngoscopy.
Conclusion
Rate of correct tracheal tube position was comparable between the three devices. Time to intubation and ventilation were shorter with MAC and Glidescope compared to VieScope. It did however show a comparable handling time to video laryngoscopy and MAC. It also did show a better visualization of the airway in the Cormack & Lehane Score compared to MAC.
Trial registration
The study was registered at the German Clinical Trials Register www.drks.de (Identifier: DRKS00024968) on March 31st 2021.
Funder
Universitätsklinikum Köln
Publisher
Springer Science and Business Media LLC
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献