Accuracy evaluation of mainstream and sidestream end-tidal carbon dioxide monitoring during noninvasive ventilation: a randomized crossover trial (MASCAT-NIV trial)

Author:

Sakuraya MasaakiORCID,Douno Eri,Iwata Wakana,Takaba Akihiro,Hadama Kosuke,Kawamura Natsuki,Maezawa ToshinoriORCID,Iwamoto Kei,Yoshino Yuya,Yoshida Kenichi

Abstract

Abstract Background The end-tidal partial pressure of carbon dioxide (PETCO2) can be used to estimate the arterial partial pressure of carbon dioxide (PaCO2) in patients who undergo mechanical ventilation via endotracheal intubation. However, no reliable method for measuring PETCO2 during noninvasive ventilation (NIV) has been established. The purpose of this study was to evaluate the correlation and agreement between PaCO2 and PETCO2 measured by these two methods and to compare them in patients who underwent NIV after extubation. Methods This study was a randomized, open-label, crossover trial in a mixed intensive care unit. We included patients who were planned for NIV after extubation and for whom the difference between PETCO2 and PaCO2 was ≤ 5 mmHg. We compared mainstream capnography using an inner cup via face mask (the novel method) with sidestream capnography (the previous method) during NIV. The relationships between PaCO2 and PETCO2 were evaluated by computing the Pearson correlation coefficient, and the agreement between PaCO2 and PETCO2 was estimated using the Bland–Altman method. Results From April 2020 to October 2021, 60 patients were included to the study. PaCO2 and PETCO2 were well correlated in both methods (the novel methods: r = 0.92, P < 0.001; the previous method: r = 0.79, P < 0.001). Mean bias between PaCO2 and PETCO2 measured using the novel method was 2.70 (95% confidence interval [CI], 2.15–3.26) mmHg with 95% limits of agreement (LoA) ranging from − 1.61 to 7.02 mmHg, similar to the result of measurement during SBT (mean bias, 2.51; 95% CI, 2.00–3.02; 95% LoA, − 1.45 to 6.47 mmHg). In contrast, measurement using the previous method demonstrated a larger difference (mean bias, 6.22; 95% CI, 5.22–7.23; 95% LoA, − 1.54 to 13.99 mmHg). Conclusion The current study demonstrated that the novel PETCO2 measurement was superior to the previous method for PaCO2 prediction. During NIV, the novel method may collect as sufficient exhalation sample as during intubation. Continuous PETCO2 measurement combined with peripheral oxygen saturation monitoring is expected to be useful for early recognition of respiratory failure among high-risk patients after extubation. Trial registration UMIN-CTR UMIN000039459. Registered February 11, 2020.

Publisher

Springer Science and Business Media LLC

Subject

Critical Care and Intensive Care Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3