Abstract
Abstract
Background
To investigate the effect of computed tomography (CT) contrast enhancement (CE) on the 3D dose distributions of non-coplanar small field beams in the CyberKnife (CK) treatment planning system (TPS) for the stereotactic ablative radiotherapy (SABR).
Methods
Twenty-two pre-CE CT treatment plans were recruited to this retrospective plan study. Their post-CE CT plans were based on the pre-CE CT plan data and calculated using the same MU and beam paths in either Ray-Tracing or Monte Carlo (MC) algorithms. The differences in the doses of the beam path and the reference point between the pre- and post-CE CT plans were compared. The minimum, maximum, and mean doses in dose-volume histograms (DVHs) of target and organs-at-risk (OARs) were also compared.
Results
The dose differences between the pre- and post-CE plans in a single beam path were less than 1.05% in both calculation algorithms, with respect to the prescription dose. At the center of the target volume, it was 1.9% (maximum 6.2%) in Ray-Tracing and 1.6% (maximum 4.0%) in MC. The CA effect showed on average 1.2% difference in the OAR maximum dose (maximum 7.8% in Ray-Tracing and 7.2% in MC). In the lung cases, the CT CE resulted in a dose difference of 2.4% (from 1.0% to 6.5%) without the calculation algorithm effect (maximum 20.3%).
Conclusions
The CK treatment plan using the post-CE CT generally afforded less than 2% dose differences from the pre-CE CT plan. However, it could be up to 7.8% depending on the target positions in a body and be more than 20% with the calculation algorithms. Thus, the post-CE CT in CK treatment plans should be used with careful consideration for the CA effect, target position, and calculation algorithm factors.
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Oncology
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献