Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery System

Author:

Armoogum Kris S,Parry John M,Souliman Salam K,Sutton David G,Mackay Colin D

Abstract

Abstract Background Intraoperative Radiotherapy (IORT) is a method by which a critical radiation dose is delivered to the tumour bed immediately after surgical excision. It is being investigated whether a single high dose of radiation will impart the same clinical benefit as a standard course of external beam therapy. Our centre has four Photon Radiosurgery Systems (PRS) currently used to irradiate breast and neurological sites. Materials and methods The PRS comprises an x-ray generator, control console, quality assurance tools and a mobile gantry. We investigated the dosimetric characteristics of each source and its performance stability over a period of time. We investigated half value layer, output diminution factor, internal radiation monitor (IRM) reproducibility and depth-doses in water. The half value layer was determined in air by the broad beam method, using high purity aluminium attenuators. To quantify beam hardening at clinical depths, solid water attenuators of 5 and 10 mm were placed between the x-ray probe and attenuators. The ion chamber current was monitored over 30 minutes to deduce an output diminution factor. IRM reproducibility was investigated under various exposures. Depth-dose curves in water were obtained at distances up to 35 mm from the probe. Results The mean energies for the beam attenuated by 5 and 10 mm of solid water were derived from ICRU Report 17 and found to be 18 and 24 keV. The average output level over a period of 30 minutes was found to be 99.12%. The average difference between the preset IRM limit and the total IRM count was less than 0.5%. For three x-ray sources, the average difference between the calculated and actual treatment times was found to be 0.62% (n = 30). The beam attenuation in water varied by approximately 1/r3. Conclusion The x-ray sources are stable over time. Most measurements were found to lie within the manufacturer's tolerances and an intercomparison of these checks suggests that the four x-ray sources have similar performance characteristics.

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging,Oncology

Reference22 articles.

1. Baum M, Vaidya JS, Mittra I: Multicentricity and recurrence of breast cancer. Lancet 1997, 349: 208. 10.1016/S0140-6736(05)60950-6

2. Vaidya JS, Baum M, Tobias JS, Morgan S, D'Souza DP: The novel technique of delivering targeted intraoperative radiotherapy (Targit) for early breast cancer. Eur J Surg Oncol 2002, 28: 447-454. 10.1053/ejso.2002.1275

3. Tobias JS, Vaidya JS, Keshtgar M, D'Souza D, Baum M: Reducing radiotherapy dose in early breast cancer: the concept of conformal intraoperative brachytherapy. Br J Radiol 2004, 77: 279-284. 10.1259/bjr/17186381

4. Vaidya JS, Tobias JS, Houghton J, Joseph D, Wenz F, Hilaris BS: Intra-operative breast radiation: the targeted intraoperative radiotherapy (Targit) trial. Breast Cancer Res Treat 2003, (Suppl 1):S2-S3.

5. Pawlik TM, Keurer HM: Accelerated partial breast irradiation as an alternative to whole breast irradiation in breast-conserving therapy for early-stage breast cancer. Women's Health 2005, 1: 59-71. 10.2217/17455057.1.1.59

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3