On the performances of Intensity Modulated Protons, RapidArc and Helical Tomotherapy for selected paediatric cases

Author:

Fogliata Antonella,Yartsev Slav,Nicolini Giorgia,Clivio Alessandro,Vanetti Eugenio,Wyttenbach Rolf,Bauman Glenn,Cozzi Luca

Abstract

Abstract Background To evaluate the performance of three different advanced treatment techniques on a group of complex paediatric cancer cases. Methods CT images and volumes of interest of five patients were used to design plans for Helical Tomotherapy (HT), RapidArc (RA) and Intensity Modulated Proton therapy (IMP). The tumour types were: extraosseous, intrathoracic Ewing Sarcoma; mediastinal Rhabdomyosarcoma; metastastis of base of skull with bone, para-nasal and left eye infiltration from Nephroblastoma of right kidney; metastatic Rhabdomyosarcoma of the anus; Wilm's tumour of the left kidney with multiple liver metastases. Cases were selected for their complexity regardless the treatment intent and stage. Prescribed doses ranged from 18 to 53.2 Gy, with four cases planned using a Simultaneous Integrated Boost strategy. Results were analysed in terms of dose distributions and dose volume histograms. Results For all patients, IMP plans lead to superior sparing of organs at risk and normal healthy tissue, where in particular the integral dose is halved with respect to photon techniques. In terms of conformity and of spillage of high doses outside targets (external index (EI)), all three techniques were comparable; CI90% ranged from 1.0 to 2.3 and EI from 0 to 5%. Concerning target homogeneity, IMP showed a variance (D5%–D95%) measured on the inner target volume (highest dose prescription) ranging from 5.9 to 13.3%, RA from 5.3 to 11.8%, and HT from 4.0 to 12.2%. The range of minimum significant dose to the same target was: (72.2%, 89.9%) for IMP, (86.7%, 94.1%) for RA, and (79.4%, 94.8%) for HT. Similarly, for maximum significant doses: (103.8%, 109.4%) for IMP, (103.2%, 107.4%) for RA, and (102.4%, 117.2%) for HT. Treatment times (beam-on time) ranged from 123 to 129 s for RA and from 146 to 387 s for HT. Conclusion Five complex pediatric cases were selected as representative examples to compare three advanced radiation delivery techniques. While differences were noted in the metrics examined, all three techniques provided satisfactory conformal avoidance and conformation.

Publisher

Springer Science and Business Media LLC

Subject

Radiology Nuclear Medicine and imaging,Oncology

Cited by 44 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3