Abstract
Abstract
Background
Therapeutic drug monitoring for voriconazole is recommended for its optimum pharmacotherapy. Although the feedback of the measurement result of serum voriconazole concentration by outsourcing needs a certain time (days within a 1 week), there was no medical equipment for the measurement available in clinical practice. Recently, a medical equipment based on high performance liquid chromatography, named LM1010, has been developed and authorized for clinical use. In this study, to validate the clinical performance of LM1010, we compared the measured serum voriconazole concentrations by LM1010 with those by outsourcing measurement using liquid chromatography-tandem mass spectrometry.
Methods
We conducted the observational study approved by the institutional review board of Kumamoto University Hospital (No. 1786). Residual serum samples harvested for therapeutic drug monitoring were separated. Measured concentrations by LM1010 by the standard filter method (needs serum volume of > 400 μL) or the dilute method (needs serum volume of 150 μL) were compared with those by outsourcing, respectively. Acceptable measurement error range of 0.72–1.33 was considered. There were 69 serum samples, where the 35 or 34 samples were employed for evaluation of the standard filter method or the dilute method, respectively.
Results
The measured concentration using the standard filter method/outsourcing was 2.22/2.10 μg/mL as the median, 1.57–3.40/1.53–3.62 as the interquartile range, < 0.2–10.76/< 0.2–11.46 μg/mL as the range, while those using the dilute method/outsourcing was 2.36/2.29 μg/mL as the median, 1.08–2.94/1.03–3.06 as the interquartile range, 0.24–10.00/< 0.2–10.85 μg/mL as the range. The regression line for the standard filter method or the dilute method were y = 0.935x + 0.154 or y = 0.933x + 0.162, respectively. The standard filter method or the dilute method showed 11.4% samples (4/35, 95%CI 3.2–26.7%) or 8.8% samples (3/34, 95%CI 1.9–23.7%) out of the acceptable measurement error range, respectively.
Conclusion
Measurement of serum voriconazole concentration by LM1010 can be acceptable in clinical TDM practice.
Funder
Japan Society for the Promotion of Science
HITACH High-Tech Science
Publisher
Springer Science and Business Media LLC
Subject
Pharmacology (medical),Pharmacology (nursing)
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献