Author:
Bell Cameron,Feizi Alborz,Roytman Gregory R.,Ramji Alim F.,Tommasini Steven M.,Wiznia Daniel H.
Abstract
Abstract
Background
Femoral head avascular necrosis (AVN), or death of femoral head tissue due to a lack of blood supply, is a leading cause of total hip replacement for non-geriatric patients. Core decompression (CD) is an effective treatment to re-establish blood flow for patients with AVN. Techniques aimed at improving its efficacy are an area of active research. We propose the use of 3D printed drill guides to accurately guide therapeutic devices for CD.
Methods
Using femur sawbones, image processing software, and 3D modeling software, we created a custom-built device with pre-determined drill trajectories and tested the feasibility of the 3D printed drill guides for CD. A fellowship trained orthopedic surgeon used the drill guide to position an 8 ga, 230 mm long decompression device in the three synthetic femurs. CT scans were taken of the sawbones with the drill guide and decompression device. CT scans were processed in the 3D modeling software. Descriptive statistics measuring the angular and needle-tip deviation were compared to the original virtually planned model.
Results
Compared to the original 3D model, the trials had a mean displacement of 1.440 ± 1.03 mm and a mean angle deviation of 1.093 ± 0.749º.
Conclusions
The drill guides were demonstrated to accurately guide the decompression device along its predetermined drill trajectory. Accuracy was assessed by comparing values to literature-reported values and considered AVN lesion size. This study demonstrates the potential use of 3D printing technology to improve the efficacy of CD techniques.
Funder
National Center for Advancing Translational Sciences
Publisher
Springer Science and Business Media LLC