Author:
Bai Xueying,Wu Tao,Zhu Yuxi,Yang Chengyu,Cheng Tiange,Liu Yi,Zhou Yi
Abstract
Abstract
Objective
To address the instability in implant surgical guides, this technique proposes an alternative anchoring mechanism in the stackable metal surgical guides utilizing cone-wedge anchors for improved stability.
Methods
Postoperative implant position superimposed onto the preoperatively planned design using Mimics Medical 21.0 and Materialise Magics 24.0 to assess 3D coronal implant deviation, 3D apical implant deviation, and implant angular deviation.
Results
Postoperative cone-beam computed tomography (CBCT) revealed a high level of precision in the implant placement, with an average 0.97 mm deviation at implant coronal region, 1.56 mm at implant apexes, and 2.95° angular deviation.
Conclusion
This technique introduces a novel cone-wedge anchoring mechanism to enhance the stability of stackable metal surgical guide templates, addressing inherent instability issues. The utilization of this approach significantly improves the accuracy of implant placement procedures.
Funder
National Natural Science Foundation of China
Publisher
Springer Science and Business Media LLC