Automated characterization of patient–ventilator interaction using surface electromyography

Author:

Sauer JuliaORCID,Graßhoff Jan,Carbon Niklas M.,Koch Willi M.,Weber-Carstens Steffen,Rostalski Philipp

Abstract

Abstract Background Characterizing patient–ventilator interaction in critically ill patients is time-consuming and requires trained staff to evaluate the behavior of the ventilated patient. Methods In this study, we recorded surface electromyography ($$\textrm{sEMG}$$ sEMG ) signals from the diaphragm and intercostal muscles and esophageal pressure ($$P_{\textrm{es}}$$ P es ) in mechanically ventilated patients with ARDS. The sEMG recordings were preprocessed, and two different algorithms (triangle algorithm and adaptive thresholding algorithm) were used to automatically detect inspiratory patient effort. Based on the detected inspirations, major asynchronies (ineffective, auto-, and double triggers and double efforts), delayed and synchronous triggers were computationally classified. Reverse triggers were not considered in this study. Subsequently, asynchrony indices were calculated. For the validation of detected efforts, two experts manually annotated inspiratory patient activity in $$P_{\textrm{es}}$$ P es , blinded toward each other, the $$\textrm{sEMG}$$ sEMG signals, and the algorithmic results. We also classified patient–ventilator interaction and calculated asynchrony indices with manually detected inspirations in $$P_{\textrm{es}}$$ P es as a reference for automated asynchrony classification and asynchrony index calculation. Results Spontaneous breathing activity was recognized in 22 out of the 36 patients included in the study. Evaluation of the accuracy of the algorithms using 3057 inspiratory efforts in $$P_{\textrm{es}}$$ P es demonstrated reliable detection performance for both methods. Across all datasets, we found a high sensitivity (triangle algorithm/adaptive thresholding algorithm: 0.93/0.97) and a high positive predictive value (0.94/0.89) against expert annotations in $$P_{\textrm{es}}$$ P es . The average delay of automatically detected inspiratory onset to the $$P_{\textrm{es}}$$ P es reference was $$-$$ - 79 ms/29 ms for the two algorithms. Our findings also indicate that automatic asynchrony index prediction is reliable. For both algorithms, we found the same deviation of $$0.06\pm 0.13$$ 0.06 ± 0.13 to the $$P_{\textrm{es}}$$ P es -based reference. Conclusions Our study demonstrates the feasibility of automating the quantification of patient–ventilator asynchrony in critically ill patients using noninvasive sEMG. This may facilitate more frequent diagnosis of asynchrony and support improving patient–ventilator interaction.

Funder

Bundesministerium für Wirtschaft und Klimaschutz

European Union – European Regional Development Fund, the Federal Government and Land Schleswig-Holstein

Drägerwerk AG & Co. KGaA

Universität zu Lübeck

Publisher

Springer Science and Business Media LLC

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3