Abstract
Abstract
Background
Catheter-based renal denervation (RDN) has been introduced to treat resistant hypertension. Although the technology of RDN has been largely improved, denervation of tortuous renal arteries remains challenging.
Case presentation
This is a case report of a 49-year-old man with drug resistant hypertension. The patient was selected for RDN after ruling out possible causes of secondary hypertension. Computed tomography angiography showed a highly tortuous left renal artery. An Iberis multielectrode ablation catheter failed to reach the target vessel with a regular guiding catheter. A 5-French extension catheter was introduced into the proximal segment of the main left renal artery to provide extra support force, which enabled successful ablation of the highly tortuous left renal artery. His ambulatory blood pressure was significantly decreased at 1 month follow-up.
Conclusions
It is feasible and effective to use a guide extension catheter for denervation of highly tortuous renal arteries. The present study provides a useful method to ablate tortuous and angled renal arteries and branches.
Funder
National Natural Science Foundation of China
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献