EnSite NavX mapping system guided implantation of a dual-chamber permanent pacemaker in a 41-year-old pregnant woman with a 4-year follow-up

Author:

Wang Peng,Wei Guang-Sheng,Wang Jun-Hua,Cao Yan-JieORCID,Zhu Wei-Wei,Shen Hang,Zhang Zhi-Yue,Ai Li,Wang Meng

Abstract

Abstract Background X-ray fluoroscopy has been the primary cardiac imaging modality in permanent pacemaker implantation (PPI) operations, but it inevitably results in radiation exposure for both operators and patients. Fluoroscopy is considered a contraindication, especially in certain circumstances, such as gestation, during which the fetus is most sensitive to radiation exposure. Therefore, measures to avoid radiation exposure are necessary, and a more safe and feasible approach is needed for this procedure. Since the EnSite NavX mapping system (ENMS) can create the required geometric contours of those relevant cardiac structures and chambers, it can be used as an alternative to X-ray fluoroscopy in PPI. In addition, because the displacement of atrial leads is a common complication of PPI, lead displacement may occur more readily without fluoroscopic guidance. Therefore, reliable measures are required to prevent leads from displacement. Case introduction A 41-year-old woman at the 15th week of gestation was referred to our department with recurrent episodes of syncope and amaurosis fugax for 2 years. Holter monitoring showed sinus rhythm, Mobitz Type II atrioventricular block and high-grade atrioventricular block with ventricular arrest up to 4945 ms. A dual-chamber PPI was performed successfully for the patient under the guidance of the ENMS instead of fluoroscopy. Displacement of atrial lead was effectively avoided by bending the top of atrial lead before implantation and making it a U-shape during operation, which left space for possible subsequent external pulling stress. Conclusions For PPI, ENMS is a feasible and reliable alternative to traditional X-ray fluoroscopy, especially when performing operations on pregnant patients. By bending the top of the active-fixation atrial lead into a U-shape during operation, the displacement of atrial lead may be avoided.

Funder

Teaching Research Project of Air Force Medical Center

Millitary Logistical Special Project for Health Care

Publisher

Springer Science and Business Media LLC

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3