Abstract
Abstract
Background
In current percutaneous coronary intervention (PCI) practice, the balloon used in the procedure should be deflated a short time after balloon dilation to avoid having prolonged balloon dilation in the coronary artery, which would block the coronary artery and cause myocardial ischemia. It is very rare for a dilated stent balloon to fail to deflate.
Case summary
A 44-year-old male was admitted to the hospital due to chest pain after exercise. Coronary angiography showed severe proximal stenosis of the right coronary artery (RCA) consistent with a diagnosis of coronary artery disease, and coronary stent implantation was required. After the last stent balloon was dilated, the stent balloon could not be deflated and continued to expand, resulting in blockage of the RCA blood flow. The patient then suffered decreased blood pressure and heart rate. Finally, the stent balloon in its expanded state was forcefully and directly withdrawn from the RCA and successfully removed from the body.
Conclusion
Deflation failure of a stent balloon is an extremely rare complication of PCI. Various treatment strategies can be considered based on hemodynamic status. In the case described herein, the balloon was pulled out of the RCA directly to restore blood flow, which kept the patient safe.
Funder
National Natural Science Foundation of China
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine
Reference13 articles.
1. Collaborators GBDCoD: Global, regional, and national age-sex-specific mortality for 282 causes of death in 195 countries and territories, 1980–2017: a systematic analysis for the Global Burden of Disease Study 2017. Lancet. 2018, 392(10159):1736–1788.
2. Mehta SR, Cannon CP, Fox KA, Wallentin L, Boden WE, Spacek R, Widimsky P, McCullough PA, Hunt D, Braunwald E, et al. Routine vs selective invasive strategies in patients with acute coronary syndromes: a collaborative meta-analysis of randomized trials. JAMA. 2005;293(23):2908–17.
3. Gruntzig A. Transluminal dilatation of coronary-artery stenosis. Lancet. 1978;1(8058):263.
4. Brown DL, Redberg RF. Last nail in the coffin for PCI in stable angina? Lancet. 2018;391(10115):3–4.
5. Ohayon P, Matta A, Boudou N. A case report of an iatrogenic coronary cameral fistula treated by retrograde percutaneous coronary intervention. Eur Heart J Case Rep. 2020;4(3):1–6.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献