Comparison of the effect of membrane sizes and fibre arrangements of two membrane oxygenators on the inflammatory response, oxygenation and decarboxylation in a rat model of extracorporeal membrane oxygenation

Author:

Edinger Fabian,Schneck EmmanuelORCID,Schulte Charlotte,Gehron Johannes,Mueller Sabrina,Sander Michael,Koch Christian

Abstract

Abstract Background Extracorporeal membrane oxygenation (ECMO) has gained widespread acceptance for the treatment of critically ill patients suffering from cardiac and/or respiratory failure. Various animal models have been developed to investigate the adverse effects induced by ECMO. Different membrane oxygenators have been used with varying priming volumes and membrane surfaces (Micro-1, small animal membrane oxygenator (SAMO)). Methods Sixteen male Lewis rats (350–400 g) were randomly assigned to receive ECMO with Micro-1 or SAMO (n = 8, respectively). Venoarterial ECMO was established after cannulation of the femoral artery and the jugular vein. The cardiac output was measured using a left-ventricular conductance catheter. The oxygen fraction of the ECMO was set to 1.0, 0.75, 0.5 and 0.21 after a stabilisation period of 15 min. Further, arterial blood gas analyses were performed at baseline, and during the first hour every 15 min after commencing the ECMO, and subsequently every 30 min. Dilutional anaemia was calculated using haemoglobin concentration at baseline, and 15 min after the start of ECMO therapy. Moreover, inflammation was determined by measuring tumour necrosis factor alpha, interleukin-6 and -10 at baseline and every 30 min. Results Animals of the Micro-1 group showed a significantly lower dilutional anaemia (ΔHaemoglobin t0 – t0.25: SAMO 6.3 [5.6–7.5] g/dl vs. Micro-1 5.6 [4.6–5.8] g/dl; p = 0.028). Further, significantly higher oxygen partial pressure was measured in the SAMO group, at an oxygen fraction of 0.75, 0.5 and 0.21 (380 [356–388] vs. 314 [263–352] mmHg, p = 0.002; 267 [249–273] mmHg vs. 197 [140–222] mmHg, p = 0.002; 87 [82–106] mmHg vs. 76 [60–79] mmHg, p = 0.021, respectively). However, no differences were found regarding the oxygen fraction of 1.0, in terms of carbon-dioxide partial pressure and cardiac output. Moreover, in the Micro-1 group tumour necrosis factor alpha was increased after 60 min and interleukin-6 after 120 min. Conclusion While the dilutional anaemia was increased after commencing the ECMO, the oxygenation was augmented in the SAMO group. The inflammatory response was elevated in the Micro-1 group.

Funder

Von-Behring-Röntgen-Stiftung

Publisher

Springer Science and Business Media LLC

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3