Author:
Benech Carlo Alberto,Perez Rosa,Benech Franco,Shirk Torrey,Bucklen Brandon S.
Abstract
Abstract
Background
Traditional minimally invasive fluoroscopy-based techniques for pedicle screw placement utilize guidance, which may require fluoroscopic shots. Computerized tomography (CT) navigation results in more accurate screw placement. Robotic surgery seeks to establish access and trajectory with greater accuracy.
Objective
This study evaluated the screw placement accuracy of a robotic platform.
Methods
Demographic data, preoperative/postoperative CT scans, and complication rates of 127 patients who underwent lumbosacral pedicle screw placement with minimally invasive navigated robotic guidance using preoperative CT were analyzed.
Results
On the GRS scale, 97.9% (711/726) of screws were graded A or B, 1.7% (12/726) of screws graded C, 0.4% (3/726) of screws graded D, and 0% graded E. Average offset from preoperative plan to final screw placement was 1.9 ± 1.5 mm from tip, 2.2 ± 1.4 mm from tail and 2.9 ± 2.3° of angulation.
Conclusions
Robotic-assisted surgery utilizing preoperative CT workflow with intraoperative fluoroscopy-based registration improves pedicle screw placement accuracy within a patient’s pedicles.
Publisher
Springer Science and Business Media LLC
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献