Author:
Yang Kai-shun,Chen Chih-Wei,Yau Ru-Bin,Liang Huang-Chien,Ko Ching-Chung,Kuo Jinn-Rung,Chio Chung-Ching,Lim Sher-Wei
Abstract
Abstract
Background
Transforaminal Lumbar Interbody Fusion (TLIF) is commonly associated with higher complications and longer operative time. This study aims to evaluate the effectiveness, safety, and usability of a novel minimally invasive surgery (MIS) bone graft delivery device.
Methods
73 consecutive patients with lumbar spondylosis, degenerative disc disease, spondylolisthesis, scoliosis or trauma were enrolled in this randomized controlled trial. Group 1 comprised 39 patients treated with the novel MIS bone graft delivery device. Group 2 consisted of 34 patients treated with the conventional system. The primary objective of the study was the assessment of the amount of bone graft delivery using the device. The secondary objectives were the effect of the device on operative time, pain relief, disability improvement, and bone fusion grade.
Results
Bone delivery amount was significantly higher in the MIS device group (6.7 ± 2.9 mL) compared to the conventional group (2.3 ± 0.5 mL), p < 0.001. Regarding the operation time, the MIS device group was associated significantly lower duration than the conventional group (p < 0.001). After a 3-month follow-up, 39.5% of the patients in the MIS device group and 3.5% of the patients in the conventional group were observed to achieve grade I fusion (complete fusion). There was a significant difference in fusion success rates (p < 0.01).
Conclusion
The novel MIS bone graft delivery device was associated with successful bone delivery. Our MIS device provides promising modality with less operative time and higher bone fusion rates than conventional modalities.
Trial Registration This trial was retrospectively registered on ClinicalTrials.gov (Registration date: 11/19/2021; Registration number: NCT05190055).
Publisher
Springer Science and Business Media LLC
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献