Background
In robot-assisted thoracic surgery (RATS) lobectomy using a robotic stapler, stapling is difficult when the stapler port place is close to the resection target vessel. We examined whether three-dimensional computed tomography (3D-CT) software enables stapler port place determination for stapling.
Methods
Seventy-three patients who underwent RATS lobectomy were enrolled. The SureForm 45 Curved-Tip stapler (136 mm from the remote center to the anvil tip) was used. The virtual distance between the resection target vessel and stapler port place (VD) was preoperatively measured with 3D-CT software. The stapler port place was the most cranial intercostal space with a VD ≥ 136 mm. The actual distance between the resection target vessel and anvil tip (AD) was measured intraoperatively. We examined the associations of the difficulty in stapling with VD, AD, chest wall damage, and clinical features.
Results
Stapling was easier with a larger anteroposterior thoracic diameter and AD. The cut-off VD and AD for smooth stapling were 142 mm and 6 mm. Chest wall damage was frequently observed at the caudal and dorsal side ports.
Conclusions
As the stapler port place is located more caudally, stapling becomes easier. However, chest wall damage increases. If the stapler port place is positioned at a site ensuring VD ≥ 142 mm by 3D-CT software, smooth stapling may be possible with a decreased incidence of chest wall damage.
Publisher
Springer Science and Business Media LLC
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Multi-Sensor Fusion Robot Positioning Algorithm based on Beidou;2023 Global Conference on Information Technologies and Communications (GCITC);2023-12-01