Author:
Nienkemper Manuel,Ludwig Björn
Abstract
Abstract
Background
Guided insertion of palatal miniscrews using a lateral cephalogram instead of cone beam computed tomography (CBCT) significantly reduces the radiation level for the patient. Till now no data are available on the risk of hitting the incisors in this regard, which is one of the worst clinical complications when inserting a paramedian miniscrew. Hence, this study aims to investigate the distance between the mini-implant and the roots of the central and lateral incisors.
Methods
Lateral cephalogram, an intraoral scan, and CBCT of 20 patients were superimposed. After a miniscrew (1.7 × 8 mm) placement based on intraoral scan and lateral cephalogram, the CBCT was used as control for the distance between the miniscrews and the roots of the incisors.
Results
The mean value of the shortest distance between the miniscrew and roots of the incisors in the lateral cephalogram was 4.74 ± 1.67 mm. The distance between both miniscrews and the central incisors measured in the CBCT was 5.03 ± 2.22 mm and 5.26 ± 2.21 mm and between the two miniscrews and the lateral incisors was 4.93 ± 1.91 mm and 5.21 ± 2.64 mm. No significant differences between the distances in the CBCT and the lateral cephalogram could be observed. In one case, the CBCT control revealed the penetration of two palatally displaced canines after insertion based on intraoral scan and lateral cephalogram.
Conclusions
The use of an intraoral scan and a lateral cephalogram for guided paramedian insertion of palatal miniscrews can prevent incisor root damage. This may reduce the radiation since no CBCT seems necessary. The current investigation focuses on the anterior paramedian area of the palate. Outside that region and in complex cases with displaced teeth in the palatal area, a CBCT might be indicated.
Funder
Universitätsklinikum Düsseldorf. Anstalt öffentlichen Rechts
Publisher
Springer Science and Business Media LLC
Subject
Neurology (clinical),General Dentistry,Otorhinolaryngology
Reference36 articles.
1. Kanomi R. Mini-implant for orthodontic anchorage. J Clin Orthod. 1997;31(11):763–7.
2. De Clerck H, Geerinckx V, Siciliano S. The Zygoma Anchorage system. J Clin Orthod. 2002;36(8):455–9.
3. Seo KW, Nahm KY, Kim SH, Chung KR, Nelson G. Chin plate with a detachable C-tube head serves for both osteotomy fixation and orthodontic anchorage. J Craniofac Surg. 2013;24(4):424–8.
4. Costa A, Raffainl M, Melsen B. Miniscrews as orthodontic anchorage: a preliminary report. Int J Adult Orthodon Orthognath Surg. 1998;13(3):201–9.
5. Melsen B, Costa A. Immediate loading of implants used for orthodontic anchorage. Clin Orthod Res. 2000;3(1):23–8.
Cited by