Author:
Zhang Miaomiao,Mao Jianqi,Ma Jia,Xu Shuqin,Lyu Yi,Yan Xiaopeng
Abstract
Abstract
Background
To investigate the feasibility of a deformable self-assembled magnetic anastomosis ring (DSAMAR), designed and developed by us, for endoscopic gastrointestinal bypass anastomosis.
Methods
Ten experimental pigs were used as model animals. The DSAMAR comprises 10 trapezoidal magnetic units, arranged in a straight line under the constraint of a guide wire. When the desired anastomosis site is reached under the guidance of an endoscope, the catheter pushes the magnetic unit along the guide wire. The linear DSAMAR can be assembled into a circular DSAMAR. Two DSAMARs were inserted, one at the end of the duodenum and the other into the stomach successively. They attracted each other and compressed the wall of the stomach and duodenum to establish gastrointestinal bypass anastomosis. The experimental pigs were euthanized 4 weeks after the operation, and the gastrointestinal bypass anastomosis specimens were obtained. The anastomosis formation was evaluated by the naked eye and histology.
Results
Gastrointestinal bypass anastomosis with DSAMARs was successfully performed. The average operation time under an endoscope was 70.30 ± 19.05 min (range: 43–95 min). The DSAMARs were discharged through the anus 10–17 days after surgery. There were no complications such as gastrointestinal bleeding, perforation, anastomotic fistula, and gastrointestinal obstruction during and after the operation. Gastroscopy and gross specimen of the anastomosis showed a well-formed magnetic anastomosis. Histological observation showed good continuity of the serous membrane and the mucosa of magnetic anastomosis.
Conclusion
The DSAMAR is a safe and feasible device for fashioning gastrointestinal bypass anastomosis in this animal model.
Funder
Fundamental Research Funds for the Central Universities
Key Research & Development Program-Social Development of Shaanxi Province of China
Magnetic surgery techniques and basic research
Institutional Foundation of The First Affiliated Hospital of Xi’an Jiaotong University
Publisher
Springer Science and Business Media LLC
Reference27 articles.
1. Lyons J, Marks J. To cut or not to cut; that is the question: surgical versus endoscopic gastrojejunostomy for the palliation of malignant gastric outlet obstruction. Gastrointest Endosc. 2023;98(3):360–1.
2. Reijm AN, Zellenrath PA, van der Bogt RD, van Driel LMJW, Siersema PD, Bruno MJ, Spaander MCW. Self-expandable duodenal metal stent placement for the palliation of gastric outlet obstruction over the past 20 years. Endoscopy. 2022;54(12):1139–46.
3. Larssen L, Hauge T, Medhus AW. Stent treatment of malignant gastric outlet obstruction: the effect on rate of gastric emptying, symptoms, and survival. Surg Endosc. 2012;26(10):2955–60.
4. Vedantam S, Shah R, Bhalla S, Kumar S, Amin S. No difference in outcomes with 15 mm vs. 20 mm lumen-apposing metal stents for endoscopic ultrasound-guided gastroenterostomy for gastric outlet obstruction: a meta-analysis. Clin Endosc. 2023;56(3):298–307.
5. Nagaraja V, Eslick GD, Cox MR. Endoscopic stenting versus operative gastrojejunostomy for malignant gastric outlet obstruction-a systematic review and meta-analysis of randomized and non-randomized trials. J Gastrointest Oncol. 2014;5(2):92–8.