Author:
Endres Kira,Marx Rudolf,Tinschert Joachim,Wirtz Dieter Christian,Stoll Christian,Riediger Dieter,Smeets Ralf
Abstract
Abstract
Background
The current surgical therapy of midfacial fractures involves internal fixation in which bone fragments are fixed in their anatomical positions with osteosynthesis plates and corresponding screws until bone healing is complete. This often causes new fractures to fragile bones while drilling pilot holes or trying to insert screws. The adhesive fixation of osteosynthesis plates using PMMA bone cement could offer a viable alternative for fixing the plates without screws. In order to achieve the adhesive bonding of bone cement to cortical bone in the viscerocranium, an amphiphilic bone bonding agent was created, analogous to the dentin bonding agents currently on the market.
Methods
The adhesive bonding strengths were measured using tension tests. For this, metal plates with 2.0 mm diameter screw holes were cemented with PMMA bone cement to cortical bovine bone samples from the femur diaphysis. The bone was conditioned with an amphiphilic bone bonding agent prior to cementing. The samples were stored for 1 to 42 days at 37 degrees C, either moist or completely submerged in an isotonic NaCl-solution, and then subjected to the tension tests.
Results
Without the bone bonding agent, the bonding strength was close to zero (0.2 MPa). Primary stability with bone bonding agent is considered to be at ca. 8 MPa. Moist storage over 42 days resulted in decreased adhesion forces of ca. 6 MPa. Wet storage resulted in relatively constant bonding strengths of ca. 8 MPa.
Conclusion
A new amphiphilic bone bonding agent was developed, which builds an optimizied interlayer between the hydrophilic bone surface and the hydrophobic PMMA bone cement and thus leads to adhesive bonding between them. Our in vitro investigations demonstrated the adhesive bonding of PMMA bone cement to cortical bone, which was also stable against hydrolysis. The newly developed adhesive fixing technique could be applied clinically when the fixation of osteosynthesis plates with screws is impossible. With the detected adhesion forces of ca. 6 to 8 MPa, it is assumed that the adhesive fixation system is able to secure bone fragments from the non-load bearing midfacial regions in their orthotopic positions until fracture consolidation is complete.
Publisher
Springer Science and Business Media LLC
Subject
Radiology Nuclear Medicine and imaging,Biomedical Engineering,General Medicine,Biomaterials,Radiological and Ultrasound Technology
Reference38 articles.
1. Endres K, Marx R, Wirtz DC, Stoll C, Riediger D, Smeets R: Adhäsive Befestigungstechnik für Osteosynthesematerial – eine In-vitro-Studie. Dtsch Zahnärztl Z 2007, 62: 317–323.
2. Luhr HG: Plattenosteosynthese in der Traumatologie des Mittelgesichtes- ein Fortschritt? Fortschr Kiefer- Gesichtschir 1991, 36: 30–33.
3. Maurer P, Bekes K, Gernhardt CR, Schaller H-G, Schubert J: Comparison of the bond strength of selected adhesive dental systems to cortical bone under in vitro conditions. Int J Oral Maxillofac Surg 2004, 33: 377–381. 10.1016/j.ijom.2003.10.020
4. Aziz SR, Ziccardi VB, Borah G: Current Therapy: Complications Associated with Rigid Internal Fixation of Facial Fractures. Compend Contin Educ Dent 2005, 8: 565–571.
5. Eppley BL, Sadove AM: Application of microfixation techniques in reconstructive maxillofacial surgery. J Oral Maxillofac Surg 1991, 49: 683–688.
Cited by
22 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献