Abstract
Abstract
Background
Some evidence has shown that malignant breast tumours have lower electrical impedance than surrounding normal tissues. Electrical impedance could be used as an indicator for breast cancer detection. The purpose of our study was to analyse the sensitivity and specificity of electrical impedance mammography (EIM) and its implementation for the differential diagnosis of pathological lesions of the breast, either alone or in combination with mammography/ultrasound, in 1200 women between 25 and 70 years old.
Methods
This study is a prospective, cross-sectional epidemiological observational study of serial screening. The women were invited to participate and signed a consent letter. Impedance imaging of the mammary gland was evaluated with the computerized mammography equipment of MEIK electroimpedance v.5.6. (0.5 mA, 50 kHz), developed and manufactured by PKF SIM-Technika®. The successful identification of breast cancer along with the sensitivity, specificity, and positive and negative predictive values of EIM were determined as follows: % sensitivity; % specificity; % positive predictive value (PPV); and % negative predictive value (NPV).
Results
EIM had a sensitivity of 85% and a specificity of 96%; the positive predictive value was 12%, and the negative predictive value was 99%. Seven cases were biopsy confirmed cancers. Significant correlations between the electrical conductivity index and body mass index (BMI) (p = 0.04) and patient age were observed (p = 0.01). We also observed that the average conductivity distribution increased according to age group (p = 0.001). We used the chi-squared test to assess the interactions between percent density and BMI (normal < 25 kg/m2 (n = 310), overweight 25–29.9 kg/m2 (n = 418) and obese ≥30 (n = 437)) (p < 0.05). The patients with a diagnosis of mammary carcinoma had a BMI of 35.51 kg/m2.
Conclusions
Our results demonstrate that the use of monofrequency electrical impedance mammography (EIM) in the detection of breast cancer had a sensitivity and specificity of 85 and 96%, respectively. These findings may support future research in the early detection of breast cancer. EIM is a non-radiation method that may also be used as a screening method for young women with dense breasts and a high risk of developing breast cancer.
Publisher
Springer Science and Business Media LLC
Subject
Cancer Research,Genetics,Oncology
Reference34 articles.
1. Cherepenin V, Karpov A, Korkenevsky A, Kornienko V, Mazaletskaya A, Mazourov D, Meister D, et al. 3D electrical impedance tomography (EIT) system for breast cancer detection. Physiol Meas. 2001;22:9–18.
2. Zain NM, Kanaga K. A review on breast electrical impedance tomography clinical accuracy. ARPN J Med Imaging. 2015;10:15.
3. Raneta O, Bella V, Bellova L, Zamecnikova E. The use of electrical impedance tomography to the differential diagnosis of pathological mammographic/sonographic findings. Neoplasma. 2013;60(6):647–54. https://doi.org/10.4149/neo_2013_083.
4. Korotkova M, Karpov A. Standars for electrical impedance mammography. In imaging of the breast. Technical aspects and clinical implications. Intech. 2012:159–78 ISBN: 978-953-51-02843-7.
5. Pak DD, Rozhkova NI, Kireeva MN, Ermoshchenkova MV, Nazarov AA, Fomin DK, Rubtsova NA. Diagnosis of Breast Cancer Using Electrical Impedance Tomography. Biomed Eng. 2012;46:154. https://doi.org/10.1007/s10527012-9292-7.
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献