Test-retest reliability and concurrent validity assessment of a novel high-frequency sensor device for anterior tibial translation measurement in loaded and unloaded condition: an exploratory cross-sectional study

Author:

Deiss Valentin,Bähler Philippe,Kolly Pascal,Schärer Anton,Henle Philippe,Eichelberger Patric,Lutz Nathanael,Baur Heiner

Abstract

Abstract Background Magnetic resonance imaging (MRI) and manual tests remain the standard for diagnosing anterior cruciate ligament (ACL) rupture. Furthermore, the passive knee displacement, also described as anterior tibial translation (ATT), is used in order to make decisions about surgery or to assess rehabilitation outcomes. Unfortunately, these manual tests are limited to passive situations, and their application to assess knee stability in loaded, weight-bearing positions are missing. Therefore, a new device with high-performance sensors and a new sensor setting was developed. The aim of this exploratory cross-sectional study was to assess the test-retest reliability of this new device in a first step and the concurrent validity in a second step. Methods A total of 20 healthy volunteers were measured. Measurement consistency of the new device was assessed on the basis of reliability during Lachman test setting and in loaded position by artificial knee perturbation in a test-retest procedure. In a second step, the concurrent validity was evaluated with the Lachmeter® as a reference instrument. Intraclass correlation coefficient (ICC), standard error of measurement (SEM), the minimal detectable change (MDC) and Bland-Altman analysis were evaluated to assess the quality criteria. Results The measurements with the new device during the Lachman test provided a mean ATT of 5.46±2.22mm. The SEM ranged from 0.60 to 0.69mm resulting in an MDC between 1.67 and 1.93mm for the new device. In the loaded test situation, the mean ATT was 2.11±1.20mm, with test-retest reliability also showing good correlation (r>0.83). The comparison of the two measurement methods with an ICC of (r>0.89) showed good correlation, which also underlines the reasonable agreement of the Bland-Altman analysis. Conclusions The evaluation of the test-retest reliability of the new device during the knee stability testing in passive situation as well as in a functional, loaded situation presented good reliability. In addition, the new device demonstrated good agreement with the reference device and therefore good validity. Furthermore, the quality criteria demonstrated the ability of the new device to detect the cut-off value (3-5mm) described in the literature for the diagnosis of ACL-deficient knees, which underlines the clinical relevance of this new device as a reliable and valid tool.

Publisher

Springer Science and Business Media LLC

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3