Comparison of accuracy for hip-knee-ankle (HKA) angle by X-ray and knee motion analysis system and the relationships between HKA and gait posture
-
Published:2023-06-03
Issue:1
Volume:24
Page:
-
ISSN:1471-2474
-
Container-title:BMC Musculoskeletal Disorders
-
language:en
-
Short-container-title:BMC Musculoskelet Disord
Author:
Zhang Hui,Chen Yanan,Jiang Huiquan,Yan Wenqing,Ouyang Yuanming,Wang Wei,Liu Yaru,Zhou Ying,Gu Shiyi,Wan Hong,He Axiang,Mao Yanjie,Liu Wanjun
Abstract
Abstract
Background
The lower limb mechanical axis was used to assess the severity of knee osteoarthritis (KOA) with varus/valgus deformity and the accuracy of targeted lower limb alignment correction after operation by conventional X-rays. There are lots of parameters to assess the gait in elder patients such as velocity, stride length, step width and swing/stance ratio by knee joint movement analysis system. However, the correlation between the lower limb mechanical axis and gait parameters is not clear. This study is aimed at obtaining the accuracy of the lower limb mechanical axis by the knee joint movement analysis system and the correlation between the lower limb mechanical axis and gait parameters.
Methods
We analysed 3D knee kinematics during ground gait of 99 patients with KOA and 80 patients 6 months after the operations with the vivo infrared navigation 3D portable knee joint movement analysis system (Opti-Knee®, Innomotion Inc, Shanghai, China). The HKA (Hip-Knee-Ankle) value was calculated and compared to X-ray findings.
Results
HKA absolute variation after the operation was 0.83 ± 3.76°, which is lower than that before the operation (5.41 ± 6.20°, p = 0.001) and also lower than the entire cohort (3.36 ± 5.72). Throughout the cohort, a significant correlation with low coefficients (r = -0.19, p = 0.01) between HKA value and anterior-posterior displacement was found. In comparing the HKA values measured on the full-length alignment radiographs and 3D knee joint movement analysis system (Opti-Knee), there was a significant correlation with moderate to high coefficients (r = 0.784 to 0.976). The linear correlation analysis showed that there was a significant correlation between the values of HKA measured by X-ray and movement analysis system (R2 = 0.90, p < 0.01).
Conclusions
Data with equivalent results as HKA, the 6DOF of the knee and ground gait data could be provided by infrared navigation based 3D portable knee joint movement analysis system comparing with the conventional X-rays. There is no significant effect of HKA on the kinematics of the partial knee joint.
Funder
Shanghai Health and Hygiene Commission
Shanghai Pudong New Area Science and Technology Development Funds
Science and Technology Commission of Shanghai Municipality
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Rheumatology
Reference31 articles.
1. Giwnewer U, Rubin G, Orbach H, Rozen N. Treat Osteoarthr Knee Harefuah. 2016;155(7):403–6.
2. Bach CM, Nogler M, Steingruber IE, Ogon M, Wimmer C, Göbel G et al. Scoring systems in total knee arthroplasty. Clin Orthop Relat Res. 2002(399):184–96. https://doi.org/10.1097/00003086-200206000-00022.
3. Oo WM, Linklater JM, Hunter DJ. Imaging in knee osteoarthritis. Curr Opin Rheumatol. 2017;29(1):86–95. https://doi.org/10.1097/bor.0000000000000350.
4. Oo WM, Bo MT. Role of Ultrasonography in knee osteoarthritis. J Clin Rheumatol. 2016;22(6):324–9. https://doi.org/10.1097/rhu.0000000000000436.
5. Stephens D. World Health Organization’s international classification of functioning, disability and health - ICF, Egs General Assembly Conference, 2001.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献