Maximum grip strength (MGS) is a reliable biomarker of overall health and physiological well-being. Therefore, an accurate and reliable measurement device is vital for ensuring the validity of the MGS assessment. This paper presents GripAble, a mobile hand grip device for the assessment of MGS. GripAble’s performance was evaluated using an inter-instrument reliability test against the widely used Jamar PLUS+ dynamometer.
Methods
MGS data from sixty-three participants (N = 63, median (IQR) age = 29.0 (29.5) years, 33 M/30 F) from both hands using GripAble and Jamar PLUS+ were collected and compared. Intraclass correlation (ICC), regression, and Bland and Altman analysis were performed to evaluate the inter-instrument reliability and relationship in MGS measurements between GripAble and Jamar PLUS+ .
Results
GripAble demonstrates good-to-excellent inter-instrument reliability to the Jamar PLUS+ with ICC3,1 = 0.906 (95% CI [0.87—0.94]). GripAble’s MGS measurement is equivalent to 69% (95% CI [0.67—0.71]%) of Jamar PLUS+’s measurement. There is a proportional difference in mean MGS between the two devices, with the difference in MGS between GripAble and Jamar PLUS+ increasing with MGS.
Conclusion
The GripAble is a reliable tool for measuring grip strength. However, the MGS readings from GripAble and Jamar PLUS+ should not be interchanged for serial measurements of the same patient, nor be translated directly from one device to the other. A new normative MGS data using GripAble will be collected and accessed through the software for immediate comparison to age and gender-matched subpopulations.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Rheumatology
Reference28 articles.
1. Sayer AA, Syddall HE, Martin HJ, Dennison EM, Roberts HC, Cooper C. Is grip strength associated with health-related quality of life? Findings from the Hertfordshire Cohort Study. Age Ageing. 2006;35(4):409–15. https://doi.org/10.1093/ageing/afl024.
2. Cui M, Zhang S, Liu Y, Gang X, Wang G. Grip strength and the risk of cognitive decline and dementia: a systematic review and meta-analysis of longitudinal cohort studies. Front Aging Neurosci. 2021;13:625551. https://doi.org/10.3389/fnagi.2021.625551 Frontiers Media S.A.
3. Leong DP, et al. Prognostic value of grip strength: Findings from the Prospective Urban Rural Epidemiology (PURE) study. The Lancet. 2015;386(9990):266–73. https://doi.org/10.1016/S0140-6736(14)62000-6.
4. Bechtol CO. Grip test; the use of a dynamometer with adjustable handle spacings. J Bone Joint Surg Am. 1954;36(4):820–32.
5. MacDermid J, Solomon G, Valdes K. and American Society of Hand Therapists; Clinical assessment recommendations 3rd edition: Impairment-based conditions. 3rd ed. New Jersey: Mount Laurel, N.J; 2015.
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献