Author:
Huang Bo-Xuan,Zhan Si-Zheng,Yang Ming,Zhang Dian-Ying
Abstract
AbstractBackgroundSelf-lock compression anti-rotation blade (SCAB) is a novel internal fixation implant for femoral neck fractures (FNF). We conducted this finite element analysis study to evaluate the biomechanical performances of SCAB combined with a cannulated screw for fixation of Pauwels type III FNF.MethodsThree finite element models of Pauwels type III FNF treated with various internal fixations were established: a: the inverted triangular parallel cannulated screw (3CS) model, b: the biplane double-supported screw fixation (BDSF) model, c: the SCAB combined with a cannulated screw model. Displacement and Von Mises stress of femurs and internal fixations under increasing loads as well as the average stress on fracture surfaces and maximum displacements on the X and Z axis of proximal fracture fragments at maximum load were measured and compared.ResultThe SCAB-based internal fixation exhibited superior biomechanical performances compared with 3CS and BDSF configurations, as the former resulted in lower parameters including displacement of the femur, Von Mises stress of internal fixation, stress on fracture surfaces as well as X and Z axis displacement of fracture fragments.ConclusionInternal fixation using SCAB combined with a cannulated screw for Pauwels type III FNFs shows enough stability, with satisfied resistance to varus and shearing forces, which may provide a new option for the treatment of FNFs.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Rheumatology
Cited by
1 articles.
订阅此论文施引文献