Finite element comparative analysis of three different internal fixation methods in the treatment of Pauwels type III femoral neck fractures

Author:

Ma Ji,Zhao Ziying,Zhi Xiaodong,Wang Hao,Wang Wei

Abstract

Abstract Background Comparison of 4 cannulated lag screws (3 inverted triangular cannulated screws + anti-rotating screws;4 CLS), dynamic hip screws + derotational screws (DHS + DS), and femoral neck fixation system (FNS) in the treatment of Biomechanical properties of middle-aged Pauwels type III femoral neck fractures. Methods The femur CT data of a healthy young volunteer was selected and imported into Mimics software to construct a three-dimensional model of a normal femur. Pauwels type III femoral neck fractures were simulated according to the 70° fracture line. Use Geomagic and SolidWorks software to optimize and build CLS, DHS + DS, and FNS fracture internal fixation models. Finally, Ansys software was used to analyze the stress distribution, peak value, and maximum displacement of the proximal fracture fragment and internal fixation; the displacement distribution, and peak value of the fracture surface at the fracture end. Results ① The stress peaks of the proximal fracture fragments in the three groups were concentrated near the femoral calcar. The peak stress of the FNS group was the largest, and the DHS + DS group was the smallest. ②The displacement of the fracture fragments was all located at the top of the femur. The peak displacement of the FNS group was the largest, and the DHS + DS group was the smallest. ③ The internal fixation stress of the three groups is concentrated in the middle part of the device. The stress distribution of the first two groups of models is more uniform than that of FNS. The peak stress of FNS is the largest and the CLS is the smallest. ④ The internal fixed displacements are all located at the top of the model. The peak displacement of the CLS is the largest, and the DHS + DS is the smallest. ⑤ The displacement of the fracture surface is in the upper part of the fractured end. The peak displacement of the FNS group was the largest, and the DHS + DS group was the smallest. Conclusion Compared with the other two internal fixation methods, dynamic hip screw + derotational screw (DHS + DS) showed good biomechanical stability. When Pauwels type III femoral neck fracture occurs in young adults, DHS + DS can be given priority as the preferred treatment for this type of fracture.

Publisher

Springer Science and Business Media LLC

Subject

Orthopedics and Sports Medicine,Rheumatology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3