Author:
Ding Zhentao,Wang Jun,Wang Yanhua,Zhang Xiaomeng,Huan Yong,Zhang Dianying
Abstract
Abstract
Background
Short-stem hip arthroplasty (SHA) is characterized by metaphyseal load transfer that effectively preserves the bone stock, but still suffers from stress shielding in the proximal femur. We designed a tension screw to mimic tension trabeculae in the new bionic collum femoris preserving (BCFP) short stem for bionic reconstruction, aiming to restore the biomechanics of hip joint.
Methods
Native femur finite element model was constructed to investigate the biomechanics of hip joint based on computed tomography (CT) data. The maximum absolute principal stress/strain cloud chart allowed the direction of stress/strain to be assessed. Six BCFP models with different screw angles (5°, 10°, 15°, 20°, 25°, and 30°) and the Corail model were created. The stress/strain distribution and overall stiffness were compared between each of the BCFP and Corail implanted models.
Results
The native model visualized the transfer pathways of tensile and compressive stress. The BCFP stems showed significantly higher stress and strain distribution in the greater trochanteric region compared to conventional total hip arthroplasty (THA). In particular, the BCFP-5° stem demonstrated the highest average strain in both medial and lateral regions and the overall stiffness was closest to the intact femur.
Conclusions
Stress transfer pathways of trabecular architecture provide biomechanical insight that serves as the basis for bionic reconstruction. The tension screw improves load transfer pattern in the proximal femur and prevents stress reduction in the greater trochanteric region. The BCFP-5° stem minimizes the stress shielding effect and presents a more bionic mechanical performance.
Funder
Peking University People’s Hospital Scientific Research Development Funds
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Rheumatology