Author:
Scott David F.,Eppich Kade,Mehić Edin,Gray Celeste,Smith Crystal Lederhos,Johnston Michael
Abstract
Abstract
Background
Hip offset, version, and length are interdependent femoral variables which determine stability and leg length. Balancing these competing variables remains a core challenge in hip arthroplasty. The potential benefits of modular femoral stems have been overshadowed by higher rates of failure. The objective of this study was to assess the survivorship of a unique dual-modular femoral stem at an average 15-year follow-up period.
Methods
The records of all patients with osteoarthritis who underwent primary total hip arthroplasty with this device between 2004–2009 were reviewed. There were no exclusions for BMI or other factors. We examined the data with Kaplan–Meier survival analysis. The primary endpoint for survival was mechanical failure of the modular neck-body junction.
Results
The survivorship of this device in 172 subjects was 100% with none experiencing mechanical failure of the modular junction at an average of 15 years. 60 patients died of causes unrelated to their THA and 9 patients were lost to follow-up. There were three early (≤ 12 months) dislocations (1.7%), and seven total dislocations (4.1%). 16 patients underwent reoperations during the follow-up period, none for any complication of the modular junction. Radiographic results showed well-fixed femoral stems in all cases. There were no leg length discrepancies of greater than 10 mm, and 85% were within 5 mm.
Conclusion
There were no mechanical failures of the modular junction in any of the subjects over the average 15-year period, demonstrating that this dual-modular design is not associated with increased failure rates. We achieved a 1.7% early dislocation rate and a 4.1% total dislocation rate without any clinically significant leg length discrepancies.
Funder
OMNIlife, Inc., Raynham, MA, USA
Publisher
Springer Science and Business Media LLC
Reference68 articles.
1. Sloan M, Premkumar A, Sheth NP. Projected Volume of Primary Total Joint Arthroplasty in the U.S., 2014 to 2030. J Bone Joint Surg. 2018;100(17):1455.
2. Klug A, Gramlich Y, Rudert M, et al. The projected volume of primary and revision total knee arthroplasty will place an immense burden on future health care systems over the next 30 years. Knee Surg Sports Traumatol Arthrosc. 2021;29(10):3287–98. https://doi.org/10.1007/s00167-020-06154-7.
3. Hariri S, Chun S, Cowan JB, Bragdon C, Malchau H, Rubash HE. Range of motion in a modular femoral stem system with a variety of neck options. J Arthroplasty. 2013;28(9):1625.
4. Sculco PK, Austin MS, Lavernia CJ, Rosenberg AG, Sierra RJ. Preventing Leg Length Discrepancy and Instability After Total Hip Arthroplasty. Instr Course Lect. 2016;65:225.
5. Turley GA, Griffin DR, Williams MA. Effect of femoral neck modularity upon the prosthetic range of motion in total hip arthroplasty. Med Biol Eng Compu. 2014;52(8):685.