Finite element analyses of lateral condyle fracture fixation in paediatrics regarding configuration of Kirschner-wire

Author:

Jeon Sangbin,Ahn Wooyeol,Oh Jongbeom,Chung Jaiwoo,Choi Junwon,Lee Soonchul

Abstract

Abstract Background This study aimed to discover the most stable outcome among different Kirschner-wire (K-wire) configurations for fixation of a lateral condyle fracture (Milch type II) in different loads of stress by using finite element analyses (FEA). Methods The right humerus of a 6-year-old boy with a lateral condyle fracture (Milch type II), was modelled with a computer aided engineering. Using FEA, peak von Mises stress and stiffness were evaluated first for a single K-wire fixation by varying the angle (0, 5, 10, 15, 20, 25, 30 degrees). Then, based on the single K-wire result, assessment of peak von Mises stress and stiffness were evaluated via FEA for two- or three-wire fixation under various configurations (two convergent, two parallel, three divergent). Results Single K-wire fixation by 5 and 25 degrees had the lowest peak von Mises stress. The fracture site showed higher stiffness at 0, 5 and 15 degrees. Considering the collected results and clinical situation, 5 degree K-wire was selected for the FEA of multiple K-wire fixation. For multiple K-wire fixation, three divergent (5–20-35 degrees) K-wires showed better stability, both in peak von Mises stress and stiffness, than any two-K-wire configurations. Among two K-wire fixations, two divergent (5–50 degrees) K-wires provided the lowest von Mises stress in varus and valgus while two divergent (5–65 degrees) K-wires showed better results in flexion, extension, internal and external rotation, and both configurations showed similar results in stiffness. Conclusions We successfully created a paediatric lateral condyle fracture (Milch type II) model which was used to conduct FEA on different K-wire configurations to achieve stability of the fracture. Our results show that an initial K-wire inserted at 5 degrees, followed by the insertion of a second divergent wire at either 45 or 60 degrees provides the most stability in two K-wire fixations in this type of fracture repair.

Funder

National research foundation, South Korea

Publisher

Springer Science and Business Media LLC

Subject

Orthopedics and Sports Medicine,Rheumatology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3