Background
To assess the impact of systematic setup and range uncertainties for robustly optimized (RO) intensity modulated proton therapy (IMPT) and volumetric modulated arc therapy (VMAT) plans in patients with localized prostate cancer.
Methods
Twenty-six localized prostate patients previously treated with VMAT (CTV to PTV expansion of 3-5 mm) were re-planned with RO-IMPT with 3 mm and 5 mm geometrical uncertainties coupled with 3% range uncertainties. Robust evaluations (RE) accounting for the geometrical uncertainties of 3 and 5 mm were evaluated for the IMPT and VMAT plans. Clinical target volume (CTV), anorectum, and bladder dose metrics were analyzed between the nominal plans and their uncertainty perturbations.
Results
With geometric uncertainties of 5 mm and accounting for potential inter-fractional perturbations, RO-IMPT provided statistically significant (p < 0.05) sparing at intermediate doses (V4000cGy) to the anorectum and bladder and high dose sparring (V8000cGy) to the bladder compared to VMAT. Decreasing the RO and RE parameters to 3 mm improved IMPT sparing over VMAT at all OAR dose levels investigated while maintaining equivalent coverage to the CTV.
Conclusions
For localized prostate treatments, if geometric uncertainties can be maintained at or below 3 mm, RO-IMPT provides clear dosimetric advantages in anorectum and bladder sparing compared to VMAT. This advantage remains even under uncertainty scenarios. As geometric uncertainties increase to 5 mm, RO-IMPT still provides dosimetric advantages, but to a smaller magnitude.