Generation of ozone during irradiation using medical linear accelerators: an experimental study

Author:

Hara N.,Oobuchi J.,Isobe A.,Sugimoto S.,Takatsu J.,Sasai K.ORCID

Abstract

Abstract Background Some patients have noted a foul odor during radiation therapy sessions, but the cause of the odor remains unknown. Since we suspected that this phenomenon is due to ozone generated by ionizing radiation, this experimental study measured ozone concentrations in the treatment room and in a coiled polyvinyl chloride (PVC) tube placed within the radiation field. Methods We measured ozone concentrations using an ultraviolet absorption method and an ozone monitor. A PVC tube (inner diameter 7 mm, outer diameter 10 mm) was used to mimic the environment of the nasal cavity. The tube (790 cm) was coiled and set between two 4-cm-thick (for X-rays) or 2-cm-thick (for electron beams) water-equivalent solid phantoms. The sampling tube of the ozone monitor was inserted into the PVC tube, and the joint was sealed to prevent environmental air contamination. To measure ozone concentrations in the atmosphere, the sampling tube supplied with the unit was used. A linac was used on a full-sized treatment field (40 cm × 40 cm at a source-to-axis distance of 100 cm). The effect of an electron beam on ozone concentrations was also evaluated with a full-sized treatment field (40 cm × 40 cm at a source-to-surface distance of 100 cm). Results Ozone levels in the treatment room were undetectable before the start of daily treatment but reached 0.008 parts per million (ppm) or more at 1 h after the start of treatment. Concentrations then remained nearly constant at 0.010–0.015 ppm throughout the day. The maximum ozone concentration in the PVC tube was only 0.006 ppm, even when it was irradiated at 2400 monitor units/min. Depending on the X-ray dose rate, the concentration increased to a maximum of 0.010 ppm with oxygen flowing into the other end of the tube at 1.5 L/min. Ozone concentrations in the PVC tube did not differ significantly between X-ray and electron-beam irradiation. Conclusions Only traces of ozone were found in the PVC tube that was used to mimic the nasal passages during radiation, these concentrations were too low for human perception. However, ozone concentrations did reach potentially detectable levels in the treatment room.

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging,Oncology

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3