Objective
To develop high-quality synthetic CT (sCT) generation method from low-dose cone-beam CT (CBCT) images by using attention-guided generative adversarial networks (AGGAN) and apply these images to dose calculations in radiotherapy.
Methods
The CBCT/planning CT images of 170 patients undergoing thoracic radiotherapy were used for training and testing. The CBCT images were scanned under a fast protocol with 50% less clinical projection frames compared with standard chest M20 protocol. Training with aligned paired images was performed using conditional adversarial networks (so-called pix2pix), and training with unpaired images was carried out with cycle-consistent adversarial networks (cycleGAN) and AGGAN, through which sCT images were generated. The image quality and Hounsfield unit (HU) value of the sCT images generated by the three neural networks were compared. The treatment plan was designed on CT and copied to sCT images to calculated dose distribution.
Results
The image quality of sCT images by all the three methods are significantly improved compared with original CBCT images. The AGGAN achieves the best image quality in the testing patients with the smallest mean absolute error (MAE, 43.5 ± 6.69), largest structural similarity (SSIM, 93.7 ± 3.88) and peak signal-to-noise ratio (PSNR, 29.5 ± 2.36). The sCT images generated by all the three methods showed superior dose calculation accuracy with higher gamma passing rates compared with original CBCT image. The AGGAN offered the highest gamma passing rates (91.4 ± 3.26) under the strictest criteria of 1 mm/1% compared with other methods. In the phantom study, the sCT images generated by AGGAN demonstrated the best image quality and the highest dose calculation accuracy.
Conclusions
High-quality sCT images were generated from low-dose thoracic CBCT images by using the proposed AGGAN through unpaired CBCT and CT images. The dose distribution could be calculated accurately based on sCT images in radiotherapy.
Funder
General Program of Jiangsu Provincial Health Commission
Changzhou Key Laboratory of Medical Physics
Changzhou Science and Technology Bureau
Young Talent Development Plan of Changzhou Health Commission
Science and Technology Programs for Young Talents of Changzhou Health Commission
Publisher
Springer Science and Business Media LLC
Subject
Radiology Nuclear Medicine and imaging,Oncology
Cited by
30 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献