Abstract
Abstract
Background
To describe the incidence, clinical characteristics, and treatment outcomes of endophthalmitis after pars plana vitrectomy (PPV) with recycled single-use devices. The recommended sterilization process as well as safety measures are discussed.
Methods
Medical charts of patients who developed endophthalmitis after PPV were retrospectively reviewed and reported in a descriptive manner. Cases undergoing PPV for preexisting endophthalmitis or open globe injury were excluded. Data collection included patient demographics, operative details, ocular findings, microbiological profiles, treatment modalities, and visual outcomes.
Results
Over the past thirteen years, a total of 12,989 pars plana vitrectomy operations were included. In total, 13 eyes of 13 cases (0.10%) experienced endophthalmitis after vitrectomy. These occurred in 3 cases (0.11%) using 20-gauge vitrectomy compared to 8 cases (0.09%) using 23-gauge vitrectomy and 2 cases (0.18%) using 25-gauge vitrectomy. There were no statistically significant differences between the 20-gauge and microincisional vitrectomy surgery (MIVS) group (P = 0.64), and the 23- and 25-gauge approach (P = 0.34). Causative pathogens were positive by culture in 5 cases (45%): 3 g-positive cases, 1 g-negative case, and 1 fungus case.
Conclusions
The rate of endophthalmitis in patients who underwent 23-gauge PPV was comparable to those who underwent 25-gauge PPV. With our standardized protocol for instrument sterilization, endophthalmitis rates in those undergoing PPV using recycled single-use instruments were within the range of previously published results in which vitrectomy tools were disposed of after one use.
Publisher
Springer Science and Business Media LLC
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献