Author:
Lombardo Luca,Palone Mario,Longo Mattia,Arveda Niki,Nacucchi Michele,De Pascalis Fabio,Spedicato Giorgio Alfredo,Siciliani Giuseppe
Abstract
Abstract
Background
To investigate and compare the gap (i.e. fit) and thickness of six aligner systems (Airnivol, ALL IN, Arc Angel, F22, Invisalign and Nuvola) using industrial computed tomography (CT). The null hypothesis was that there would be no detectable differences in either measurement between the aligners investigated.
Materials and methods
Passive aligners of each brand were fitted to one single resin cast prototyped from an STL file from a single patient. The samples obtained were examined under high-resolution micro-CT, and the resulting tomographic microphotographs and volumetric data were compared. 3D analysis investigated the gap volume, the mean gap width and the maximum gap width of each sample. A total of 204 linear 2D measurements were made on 18 microtomographic images to investigate the aligner gap and thickness among different systems. Investigated regions were the central incisor, canine and first molar. The resulting measurements were analysed by ANOVA and compared using Tukey’s post hoc analysis (P < 0.05).
Results
3D analysis revealed that the F22 displayed lower gap volume and mean gap width, followed by Airnivol and Invisalign, whereas Airnivol the lowest maximum gap width. 2D analysis showed that F22 had the lowest mean gap and aligner thickness at all teeth investigated. Comparison of the 2D point values revealed statistically significant differences between brands in terms of both measurements (P < 0.05), with the exception of six points in the gap analysis and one in the thickness analysis.
Conclusions
There are differences between the six aligner systems examined in terms of 2D and 3D measurements of aligner thickness and gap.
Publisher
Springer Science and Business Media LLC
Reference27 articles.
1. Kuo E, Miller RJ. Automated custom-manufacturing technology in orthodontics. Am J Orthod Dentofacial Orthop. 2003;123:578–81.
2. Jeremiah HG, Bister D, Newton JT. Social perceptions of adults wearing orthodontic appliances: a cross-sectional study. Eur J Orthod. 2011;33(5):476–82.
3. Rossini G, Parrini S, Castroflorio T, Deregibus A. Debernardi CL. Periodontal health during clear aligners treatment: a systematic review Eur J Orthod. 2015;37(5):539–43.
4. Rossini G, Parrini S, Castroflorio T, Deregibus A, Debernardi CL. Efficacy of clear aligners in controlling orthodontic tooth movement: a systematic review. Angle Orthod. 2015;85(5):881–9.
5. Papadimitriou A, Mousoulea S, Gkantidis N, Kloukos D. Clinical effectiveness of Invisalign® orthodontic treatment: a systematic review. Prog Orthod. 2018;28;19(1):37.
Cited by
33 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献