Abstract
Abstract
Background
The loss of third-order information in pre-adjusted brackets due to torsional play is a problem in clinical orthodontics. The aim of this study was to evaluate the impact of slot height, archwire height, width and edge bevel’s radius on the torsional play for three brackets/archwire systems.
Methods
Ninety brackets with a 0.022 × 0.028 in. slot with McLaughlin-Bennett-Trevisi prescription from three different manufacturers were selected, and the slot’s height and depth were measured using a profile projector. Sixty stainless-steel rectangular archwires from three different manufacturers were sectioned and observed with a SEM to measure their height, width, and radius of edge bevel. The recorded data were used to calculate the theoretical torsional play between different slot−archwire combinations. One-way ANOVA was used to compare the measurements within different bracket types and among different manufacturers.
Results
Slot height was usually oversized. Archwire’s height was usually undersized, but oversized wires were also observed. The radius edge bevel was the most variable parameter. A certain degree of torsional play is always present that differs from one bracket type to another of the same producer and that can even be doubled from one manufacturer to another.
Conclusions
Due to production tolerance, differences between the nominal values and the real dimensions of any components of a slot/archwire system are common. This results in a torsional play that limits torque expression. The archwire’s edge bevel plays an important role in torque expression, and clearer information should be provided by the manufacturers regarding this aspect.
Publisher
Springer Science and Business Media LLC
Reference23 articles.
1. Joch A, Pichelmayer M, Weiland F. Bracket slot and archwire dimensions: manufacturing precision and third order clearance. J Orthod. 2010;37:241–9.
2. Gioka C, Eliades T. Materials-induced variation in the torque expression of preadjusted appliances. Am J Orthod Dentofacial Orthop. 2004;125:323–8.
3. Meling TR, Ødegaard J. The effect of cross-sectional dimensional variations of square and rectangular chromecobalt archwires on torsion. Angle Orthod. 1998;68:239–48.
4. Cash AC, Good SA, Curtis RV, McDonald F. An evaluation of slot size in orthodontic brackets--are standards as expected? Angle Orthod. 2004;74:450–3.
5. Sebanc J, Brantley WA, Pincsak JJ, Conover JP. Variability of effective root torque as a function of edge bevel on orthodontic arch wires. Am J Orthod. 1984;86:43–51.
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献