Author:
Momose Haruka,Murayama Asako,Yamada Norie,Matsubayashi Keiji,Matsuoka Sahoko,Ikebe Emi,Kuramitsu Madoka,Muramatsu Masamichi,Kato Takanobu,Hamaguchi Isao
Abstract
Abstract
Background
Hepatitis B virus (HBV) infection is a global public health concern. Precise and sensitive detection of viral markers, including HBV DNA and HBs antigen (Ag), is essential to determine HBV infection.
Methods
The sensitivities and specificities of 5 HBV DNA and 14 HBsAg kits were evaluated using World Health Organization International Standards (WHO IS) and the Regional Reference Panel (RRP) consisting of 64 HBsAg-negative and 80 HBsAg-positive specimens.
Results
All 5 HBV DNA kits detected HBV DNA in the WHO IS at a concentration of 10 IU/mL. The sensitivity and specificity to the RRP were 98.8–100% and 96.9–100%, respectively. HBV DNA titers were well correlated among the 5 kits regardless of HBV genotype. However, discordance of the HBV DNA titer was found in 5 specimens measured by CAP/CTM HBV v2.0. Among 12 automated HBsAg kits, the minimum detectable concentrations in the WHO IS varied from 0.01 to 0.1 IU/mL. Two lateral flow assays were positive for WHO IS concentrations greater than or equal to 1.0 and 0.1 IU/mL, respectively. When analyzed by the RRP, 12 automated kits exhibited a sensitivity of 98.8–100%, and 2 lateral flow assays showed sensitivities of 93.8% and 100%. The specificities of HBsAg kits were 100%. In the quantification of HBsAg, some kits showed a poor correlation of measurements with each other and showed up to a 1.7-fold difference in the regression coefficient of HBsAg titers. There were variations in the correlations of measurements among HBsAg kits when analyzed by genotype.
Conclusions
Five HBV DNA kits showed sufficient sensitivity and specificity to determine HBV infection. HBV DNA titers were compatible with each other irrespective of HBV genotypes. HBsAg kits had enough sensitivity and specificity to screen for HBV infection. One of the lateral flow assays had a nearly equivalent sensitivity to that of the automated HBsAg kit. HBsAg titers quantified by the evaluated kits were not compatible across the kits. Genotype-dependent amino acid variations might affect the quantification of HBsAg titers.
Funder
Japan Agency for Medical Research and Development
Publisher
Springer Science and Business Media LLC
Subject
Infectious Diseases,Virology
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献