Abstract
Abstract
Introduction
Pseudomonas aeruginosa is a common cause of ventilator-associated pneumonia (VAP). Rapid and accurate detection of lower respiratory tract colonization and/or infection with P. aeruginosa may advise targeted preventive (antibody-based) strategies and antibiotic therapy. To investigate this, we compared semi-quantitative culture results from 80 endotracheal aspirates (ETA) collected from mechanically-ventilated patients, to two culture and two non-culture-based methods for detection of P. aeruginosa.
Methods
P. aeruginosa-positive (n = 40) and -negative (n = 40) ETAs from mechanically ventilated patients analyzed initally by (i) routine semi-quantitative culture, were further analyzed with (ii) quantitative culture on chromogenic ChromID P. aeruginosa and blood agar; (iii) enrichment in brain heart infusion broth followed by plating on blood agar and ChromID P. aeruginosa; (iv) O-antigen acetylase gene-based TaqMan qPCR; and (v) GeneXpert PA PCR assay.
Results
Of the 80 ETA samples included, one sample that was negative for P. aeruginosa by semi-quantitative culture was found to be positive by the other four methods, and was included in an “extended” gold standard panel. Based on this extended gold standard, both semi-quantitative culture and the GeneXpert PA assay showed 97.6% sensitivity and 100% specificity. The quantitative culture, enrichment culture and O-antigen acetylase gene-based TaqMan qPCR had a sensitivity of 97.6%, 89.5%, 92.7%, and a specificity of 97.4%, 100%, and 71.1%, respectively.
Conclusion
This first evaluation of the GeneXpert PA assay with ETA samples found it to be as sensitive and specific as the routine, hospital-based semi-quantitative culture method. Additionally, the GeneXpert PA assay is easy to perform (hands-on time ≈ 5 min) and rapid (≈ 55 min assay time). The combination of the high sensitivity and high specificity together with the rapid acquisition of results makes the GeneXpert PA assay a highly recommended screening technique. Where this equipment is not available, semi-quantitative culture remains the most sensitive of the culture methods evaluated here for P. aeruginosa detection in ETA samples.
Funder
Innovative Medicines Initiative
Publisher
Springer Science and Business Media LLC
Subject
Pharmacology (medical),Infectious Diseases,Microbiology (medical),Public Health, Environmental and Occupational Health
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献