Customized-individually-made origin® implants in total knee arthroplasty allow a reliable solution for accurate reproduction of planned implant positioning

Author:

Sajan Martin,Moussa Mohamad K.,Lefèvre Nicolas,Payan Charles,Valentin Eugénie,Meyer Alain,Bohu Yoann,Khalaf Zeinab,Grimaud Olivier,Gerometta Antoinne,Hardy Alexandre,Khiami Frédéric

Abstract

Abstract Purpose To evaluate the accuracy and reproducibility of a patient-specific, customized individually made (CIM) total knee replacement (TKR) using the ORIGIN® prosthesis. Methods This was a prospective study conducted at a University Hospital from January 15, 2019, to April 30, 2021. The study included patients planned for an ORIGIN® CIM TKR procedure. Exclusion criteria included revision surgery, severe deformity, stiffness, or laxity. Evaluations were carried out using computed tomography scans performed 8 weeks preoperatively and 6 weeks postoperatively. The primary outcome measurements were the preoperative, planned, and postoperative CT scan alignment measurements including the Hip-Knee-Ankle (HKA) angle, mechanical Medial Distal Femoral articular surface Angle (mMDFA, distal alpha angle), Posterior Distal femoral articular surface angle (PDFA, posterior alpha angle), mechanical Medial Proximal Tibial articular surface Angle (mMPTA, beta angle) and posterior proximal tibial angle (PPTA). Secondary outcomes included the accuracy of implant positioning with percentage of outliers at 2° and 3° Results The study encompassed 51 knees from 50 patients with mean age of 68.1 (SD = 8.89). The overall HKA angle deviated by -0.93° [95% CI: -1.45; -0.43], and the PDFA angle by -0.61° [95% CI: -1.07; -0.15], while the mMPTA exceeded planned values by 1.00° [95% CI: 0.57; 1.43]. The 3° outliers rate ranged from 3.9% for the mMPTA to 7.8% for the HKA alignment, with no outliers in mMDFA and PPTA. Similarly, the 2° outliers rate ranged from 15.7% for both the PDFA angle and mMPTA to 19.6% for the HKA alignment. The Bland–Altman plots further emphasized the precision of planned and post-operative angles across all measurements. Conclusion The CIM TKR showed high accuracy and reproducibility, closely matching preoperative planning. The weakest accuracy at 3°-outliers is in the reproduction of the HKA alignment at 92.2% (range for all angle: 92.2–100%). Similarly, the weakest accuracy at 2°-outliers is in the reproduction of the HKA alignment at 80.4% (range for all angles: 80.4–92.2%).

Publisher

Wiley

Subject

Orthopedics and Sports Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3