Abstract
Abstract
Purpose
Soft-tissue tension around the hip joint is related to the incidence of dislocation after total hip arthroplasty (THA), but it remains difficult to quantify the soft-tissue tension during surgery. In this study, a three-dimensional force sensor-instrumented modular femoral head was developed and used to quantify soft-tissue tension during THA. The forces at the hip joint were also calculated using a three-dimensional musculoskeletal computer model to validate the measured forces.
Methods
Soft-tissue tension was investigated by measuring the hip joint forces and directions during intraoperative trialing in four patients through passive range of motion (ROM) from 0° extension to 90° flexion. A musculoskeletal model with THA, which was scaled to one of four patients, was developed. The hip joint forces were calculated under the same motion.
Results
Through the passive ROM, the magnitude of soft-tissue tension was greatest when the hip was extended, decreased with flexion to 34°, and progressively increased to flexion at 90°. The mediolateral force component was relatively constant, but the supero-inferior and anterior–posterior force components changed significantly. Within-individual variations were small during three repeated cycles of measurement, but magnitudes varied significantly among patients. Similar force patterns and magnitudes were calculated by the musculoskeletal model.
Conclusions
This study demonstrates that it is possible to quantify soft-tissue tension and direction during THA with an instrumented head. There was general agreement between the calculated and measured forces in both pattern and magnitude. Including additional subject-specific details would further enhance agreement between the model and measured hip forces.
Subject
Orthopedics and Sports Medicine
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献