Abstract
Abstract
Background
Inertial measurement unit (IMU)-based motion capture systems are gaining popularity for gait analysis outside laboratories. It is important to determine the performance of such systems in specific patient populations. We aimed to validate and determine within-day reliability of an IMU system for measuring lower limb gait kinematics and temporal–spatial parameters (TSP) in people with and without HIV.
Methods
Gait was recorded in eight adults with HIV (PLHIV) and eight HIV-seronegative participants (SNP), using IMUs and optical motion capture (OMC) simultaneously. Participants performed six gait trials. Fifteen TSP and 28 kinematic angles were extracted. Intraclass correlations (ICC), root-mean-square error (RMSE), mean absolute percentage error and Bland–Altman analyses were used to assess concurrent validity of the IMU system (relative to OMC) separately in PLHIV and SNP. IMU reliability was assessed during within-session retest of trials. ICCs were used to assess relative reliability. Standard error of measurement (SEM) and percentage SEM were used to assess absolute reliability.
Results
Between-system TSP differences demonstrated acceptable-to-excellent ICCs (0.71–0.99), except for double support time and temporophasic parameters (< 0.60). All TSP demonstrated good mean absolute percentage errors (≤7.40%). For kinematics, ICCs were acceptable to excellent (0.75–1.00) for all but three range of motion (ROM) and four discrete angles. RMSE and bias were 0.0°–4.7° for all but two ROM and 10 discrete angles. In both groups, TSP reliability was acceptable to excellent for relative (ICC 0.75–0.99) (except for one temporal and two temporophasic parameters) and absolute (%SEM 1.58–15.23) values. Reliability trends of IMU-measured kinematics were similar between groups and demonstrated acceptable-to-excellent relative reliability (ICC 0.76–0.99) and clinically acceptable absolute reliability (SEM 0.7°–4.4°) for all but two and three discrete angles, respectively. Both systems demonstrated similar magnitude and directional trends for differences when comparing the gait of PLHIV with that of SNP.
Conclusions
IMU-based gait analysis is valid and reliable when applied in PLHIV; demonstrating a sufficiently low precision error to be used for clinical interpretation (< 5° for most kinematics; < 20% for TSP). IMU-based gait analysis is sensitive to subtle gait deviations that may occur in PLHIV.
Funder
South African Medical Research Council
Harry Crossley Foundation
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Biomedical Engineering,General Medicine,Biomaterials,Radiological and Ultrasound Technology
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献