Author:
Nuzul Sigit Mohammad,Kartinah Neng Tine,Nur Busjra M.,Ilyas Ermita I. Ibrahim
Abstract
Abstract
Background
Tube misplacement into the tracheobronchial tract is associated with pneumothorax in 0.5% of cases. NGT verification only detects the position of the tube at the end of the procedure. Therefore, a safe nasogastric tube (SNGT) was created to detect the NGT position in real time in a simple and inexpensive way. This study aimed to prove the effectiveness of the SNGT prototype in Macaca fascicularis.
Result
An SNGT producing 50% of the TV had 100% sensitivity and specificity in detecting the position of the tube at 100% of the TV, with a sensitivity of 100% and a specificity of 87.5%. There was a significant difference between the movement of the SNGT 50% TV and SNGT 100% TV airbags (p ≤ 0.05). However, there was no significant difference between the accuracy of placement of the 50% TV SNGT, 100% TV SNGT, and conventional NGT (p > 0.05). The pepsin enzyme had better sensitivity (100%) than pH paper (91.66%) in detecting the end-of-procedure tube position. This research has the potential to advance into human clinical trials.
Conclusion
SNGTs are highly effective in detecting the NGT position inside the respiratory and digestive tracts to prevent misplacement.
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging,Biomedical Engineering,General Medicine,Biomaterials,Radiological and Ultrasound Technology