Abstract
Abstract
Background
Unretrievable foreign bodies are associated with high morbidity and mortality. While the majority of reported cases involve the venous circulation, intra-arterial foreign body displacement have the potential to migrate more distally with a higher risk for dissection and hemorrhagic complications during retrieval. As the number of intravascular procedures continues to increase, there is also likely to be a concomittant increase in the number of retrieval procedures, particular for fractured catheters and sheaths. Although snaring is frequently the traditional, ‘go-to’ method for retrieval, there are inherent risks of further dislodgement or fracture.
Case report
We describe a case that involves retrieval of a fractured sheath that originated in the common femoral artery but migrated into the popliteal artery. Different retrieval approaches were employed, however ultimately balloon assisted, over-the-wire retrieval was the successful approach.
Conclusions
We anticipate that over-the-wire, inline-retrieval approaches will continue to grow in popularity and use, particularly with respect to manipulation within the arterial circulation.
Funder
National Center for Advancing Translational Sciences
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Radiology Nuclear Medicine and imaging
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献