Abstract
Abstract
Background
Current guidelines recommend the use of mechanical circulatory support (MCS) for patients with cardiogenic shock that is refractory to medical therapy. Bleeding is the most common complication of MCS. Transarterial embolization (TAE) is often performed to treat this complication, because it is a less invasive hemostatic procedure. However, the TAE option needs to be carefully considered during MCS, as the access route may be limited during MCS.
Case presentation
A man in his 70 s was diagnosed with acute myocardial infarction and underwent percutaneous coronary intervention via venoarterial extracorporeal membrane oxygenation (VA-ECMO) and Impella. During treatment in the intensive care unit, he suffered damage to a branch of the internal thoracic artery during a cardiac drainage procedure, which was subsequently treated via emergency TAE.
An ECMO return cannula and an Impella sheath were inserted into the patient’s right and left femoral arteries, respectively. An approach from the left brachial artery was selected, and the left internal thoracic artery was embolized. Subsequently, the patient required re-intervention to treat re-bleeding from another artery. Because it was difficult to target the target artery from the brachial one, owing to interference from the Impella catheter, the ECMO circuit near the return cannula was punctured and a guiding sheath was inserted. The ECMO flow and the patient’s blood pressure decreased following placement of this guiding sheath. We were thus able to maintain the patient’s blood pressure by increasing the infusion fluids and Impella flow, and embolize the target artery using a gelatin sponge to achieve hemostasis.
Conclusion
When TAE is difficult to perform during MCS using an approach from the upper extremities, a lower extremity approach with a sheath inserted into the ECMO circuit may represent a viable alternative.
Publisher
Springer Science and Business Media LLC