The present study attempted to demonstrate how the configuration of hydrogel-coated coils is influenced by different temperature exposures.
Thirty detachable hydrogel-coated coils were evaluated in an in vitro water immersion test under five different temperature ranges (22.6 °C, 37 °C, 40–50 °C, 50–60 °C, and 60–70 °C). The configuration changes were classified (configuration I, configuration II, and configuration III) according to the curling that occurred during 30 min of immersion. Configuration stability of five Hydrogel-coated coils was also evaluated in a two-step temperature immersion test.
Results
All hydrogel-coated coils showed some configuration changes during water immersion. However, a logarithmic transformation of the time and temperature data showed a significant (p < 0.05) negative linear correlation between time and temperature for all coil configurations (configuration I: R = 0.97, configuration II: R = 0.98, configuration III: R = 0.97). The time needed to reach configuration III (complete coiling) was 160.4 ± 41.9 s at 37.5 °C (range: 100–205 s), 45.7 ± 22.2 s at 47.5 °C (range: 23–70 s), 20.2 ± 7.2 s at 57.5 °C (range: 14–32 s), and 10.3 ± 2.4 s at 67.5 °C (range: 7–13 s).
Conclusions
Temperatures above 55 °C induced immediate configurational changes in the hydro-coated coils, achieving complete curling within less than 30 s. Temperatures near 36 °C (normal body temperature) require more time to reach optimal coil curling (configuration III). The optimization of HydroCoil preparation can reduce interventional procedural time and improve clinical results.
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging