Abstract
Abstract
Background/purpose
To study steerable microcatheter (SM) use in moderate and highly difficult vessel selection compared to conventional pre-shaped microcatheter (CM) use.
Material and methods
An IRB approved, single institution analysis of 40 complex angiographic procedures with and without superselective microcatheter use during an eight-month period in 2017 was performed. Target vessels were deemed moderate or highly difficult to select based on vessel size, tortuosity, and/or angulation during non-selective initial angiography. Data collected included type of microcatheter used (SM or CM), number of microcatheters and microwires used, procedure time, radiation exposure index (dose area product/DAP), target vessel location, and time to target vessel selection (TTVS; time from device placement to vessel selection). Comparison between the SM and CM groups was performed using Wilcoxon test.
Results
A SM (SwiftNinja, Merit Medical, South Jordan, UT, USA) was used to select 46 vessels in 20 patients. One or more CMs were used in 20 patients to select 34 vessels. Median TTVS, number of microwires used, total procedure time, and DAP (microGray.m2) were 12 vs. 462.5 s (p < 0.0001), 0 vs. 2 (p < 0.001), and 26,948 vs. 30,904 (p = 0.15) in the SM vs. CM groups, respectively. When adjusted for body mass index (BMI) using a linear model for radiation exposure, patients in the SM group had lower radiation exposure than those in the CM group (p = 0.05).
Conclusions
Utilization of a steerable microcatheter, without or with a guidewire, leads to easier and faster target vessel selection with shorter procedure times in complex vessel anatomy.
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Radiology Nuclear Medicine and imaging
Reference13 articles.
1. Bhatia S, Sinha VK, Abdul-Rahim O, Harward S, Narayanan G (2018) Rare prostatic artery origins and the importance of collateral circulation in prostate artery embolization: a pictorial essay. Can Assoc Radiol J 69(2):220–229
2. Carnevale FC, Moreira AM, Antunes AA (2014) The “PErFecTED technique”: proximal embolization first, then embolize distal for benign prostatic hyperplasia. Cardiovasc Intervent Radiol 37(6):1602–1605
3. Ector J, Dragusin O, Adriaenssens B, Huybrechts W, Willems R, Ector H, Heidbuchel H (2007) Obesity is a major determinant of radiation dose in patients undergoing pulmonary vein isolation for atrial fibrillation. J Am Coll Cardiol 50(3):234–242
4. Er F, Yuksel D, Hellmich M, Gassanov G (2015) Comparison of conventional versus steerable-catheter guided coronary sinus Lead positioning in patients undergoing cardiac resynchronization device implantation. PLoS One 10(11):e0143292
5. Galbraith AS, Wallace E, Devitt A (2018) Examining the association of body mass index and the depth of epidural space, radiation dose exposure and fluoroscopic screening time during transforaminal nerve block injection: a retrospective cohort study. Ir J Med Sci 188:1–8
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献