Altit Gabriel,Bhombal Shazia,Chock Valerie Y.
Abstract
Abstract
Background
Near-infrared spectroscopy (NIRS) measures of cerebral saturation (Csat) and renal saturation (Rsat) in extreme premature newborns may be affected by systemic blood flow fluctuations. Despite increasing clinical use of NIRS to monitor tissue saturation in the premature infant, validation of NIRS measures as a correlate of blood flow is still needed. We compared echocardiography (ECHO) derived markers of ascending aorta (AscAo) and descending aorta (DesAo) blood flow with NIRS measurements obtained during the ECHO.
Methods
Newborns < 29 weeks’ gestation (2013–2017) underwent routine NIRS monitoring. Csat, Rsat and systemic saturation at the time of ECHO were retrospectively analyzed and compared with Doppler markers of aortic flow. Renal and cerebral fractional tissue oxygen extraction (rFTOE and cFTOE, respectively) were calculated. Mixed effects models evaluated the association between NIRS and Doppler markers.
Results
Forty-nine neonates with 75 Csat-ECHO and 62 Rsat-ECHO observations were studied. Mean post-menstrual age was 28.3 ± 3.8 weeks during the ECHO. Preductal measures including AscAo velocity time integral (VTI) and AscAo output were correlated with Csat or cFTOE, while postductal measures including DesAo VTI, DesAo peak systolic velocity, and estimated DesAo output were more closely correlated with Rsat or rFTOE.
Conclusions
NIRS measures are associated with aortic blood flow measurements by ECHO in the extremely premature population. NIRS is a tool to consider when following end organ perfusion in the preterm infant.
Publisher
Springer Science and Business Media LLC
Subject
Pediatrics, Perinatology, and Child Health
Cited by
4 articles.